GE Healthcare MAC 5

Error: Image Is Not Found in the USB Path

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

GE Healthcare

Model

MAC 5

What This Guide Helps With

This guide helps troubleshoot a GE Healthcare MAC 5 USB software upgrade failure when the device displays Error: Image Is Not Found in the USB Path. This condition means the required software image is not present on the inserted USB flash drive, is not in the expected location, or cannot be read by the device during the USB upgrade process.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Confirm the USB Drive Is Inserted and Detected

Check the USB Drive Contents on a Computer

Verify the Software Image Is in the Correct USB Location

Confirm the Correct Software Version and Package

Confirm the Digital Signature File Is Present if Required

Check for File Copy or Download Problems

Try a Known-Good USB Flash Drive

Check USB Formatting and Access

Repeat the USB Software Upgrade Attempt

Stop When Resolved

If the Problem Persists

If the correct digitally signed MAC 5 software image is confirmed to be present in the correct USB path, the USB drive is known good, the file location is correct, and the error still returns, external and common causes have been ruled out.

At that point, the issue may involve USB port failure, storage access problems, software upgrade process failure, or another device-side fault.

Remove the MAC 5 from service, label it Out of Service, and send it for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot software upgrade errors on an active patient or while the device is needed for patient ECG acquisition. Move the patient or testing workflow to a backup ECG machine first, maintain therapy and diagnostic continuity, and only return the MAC 5 to service after the update issue is resolved and functional verification is complete.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"GE Healthcare MAC 5 displayed Error: Image Is Not Found in the USB Path during USB software upgrade. Device could not locate the required software image on the inserted USB flash drive."

Cause

What was observed during troubleshooting.

Example:
"Software image was not present in the required USB path, incorrect package was loaded, USB media was unreadable, or device-side USB/software access issue remained after external checks."

Resolution

What action was taken.

Example:
"Verified USB detection, checked USB contents, confirmed correct digitally signed MAC 5 software image and file location, recopied upgrade files to known-good USB drive, and retried upgrade. If unresolved, device was removed from service and sent for bench evaluation."

Helpful Details to Include (If Known)

Final Thought

Software upgrade errors should be handled carefully because the device depends on a valid, complete, and correctly placed signed image. Start with the USB drive, file location, and software package before assuming device failure. If the correct image is confirmed and the error persists, escalation is appropriate. Clear CCR documentation protects the device history and supports safe return to service.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?