GE Healthcare MAC 5500 HD

Date, Time, Or Network Time Synchronization Error

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

GE Healthcare

Model

MAC 5500 HD

What This Guide Helps With

Troubleshooting incorrect date or time, clock drift, or failed network synchronization caused by settings, connectivity, time-source, or internal clock problems.

Step-by-Step Troubleshooting

1. Protect Patient Record Accuracy

Do not transmit or finalize an ECG with an incorrect acquisition date or time without following the facility-approved correction process.

Notify clinical staff of the timestamp issue.

Verify the actual current date and time from an approved source.

Prevent duplicate ECGs created solely to correct a timestamp.

Document any patient studies acquired while the clock was incorrect.

Expected outcome: Incorrect timestamps are recognized before they cause patient-record, interpretation, or legal-documentation errors.

2. Confirm the Exact Clock Problem

Determine whether:

Expected outcome: The problem is defined as manual-setting error, drift, reset, time-zone issue, or network synchronization failure.

3. Compare With an Approved Time Source

Compare the MAC 5500 HD clock with the facility’s approved reference.

Record:

Expected outcome: The magnitude and pattern of the time error are documented accurately.

4. Check Network Connectivity

If the device normally receives network time:

Expected outcome: Network communication is confirmed or identified as the reason synchronization cannot occur.

5. Check Whether Multiple Devices Are Affected

Review another MAC system or network-connected clinical device in the same area.

If several devices show the same error, contact IT or the time-server administrator.

If only one MAC 5500 HD is affected, continue device-level troubleshooting.

Expected outcome: A local device problem is separated from an infrastructure or time-server problem.

6. Restart the Electrocardiograph

When no ECG is being acquired, saved, or transmitted:

Expected outcome: The clock corrects and remains stable after restart, or the error repeats consistently.

7. Verify Authorized Date, Time, and Time-Zone Settings

Qualified personnel may review normal authorized settings.

Confirm the configured time zone is appropriate.

Confirm daylight-saving configuration aligns with facility policy.

Do not alter restricted system, network, certificate, or time-source settings without authorization.

Record the original settings before any approved change.

Expected outcome: Incorrect authorized settings are corrected and the displayed time matches the approved reference.

8. Evaluate Clock Retention

After setting or synchronizing the clock:

Expected outcome: The device retains the correct clock. Failure to retain it indicates a service-level problem.

9. Verify Record Timestamps and MUSE Alignment

Use an approved test ECG.

Confirm the acquisition timestamp on the MAC 5500 HD.

Transmit the test ECG.

Confirm the timestamp shown in MUSE.

Verify device, MUSE, and reference time agree closely enough for the established workflow without inventing a tolerance.

Expected outcome: The complete ECG record displays the correct date and time locally and in MUSE.

10. Document Potentially Affected Patient Records

If ECGs were acquired while the clock was incorrect:

Expected outcome: Potential record discrepancies are communicated and managed through the proper clinical documentation process.

If the Problem Persists

Network connectivity, authorized settings, restart, time zone, and infrastructure causes have been ruled out. The remaining possibilities may include internal clock retention failure, system-board timekeeping fault, corrupted configuration, failed time synchronization service, or certificate-related communication problems.

The device should be:

Clinical Use Tip

Incorrect ECG time can affect event correlation, treatment decisions, and legal documentation even when the waveform itself is accurate.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that ECGs from the MAC 5500 HD displayed a time one hour earlier than the actual acquisition time."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found that the device time-zone setting did not match the facility’s approved configuration."

Resolution

What action was taken.

Example:
"Clinical Engineering corrected the authorized time-zone setting and verified matching timestamps on the device and a transmitted MUSE test record."

Helpful Details to Include (If Known)

Final Thought

Treat date and time as part of the clinical record, verify network and settings before assuming hardware failure, protect affected ECGs, and escalate persistent clock-retention or synchronization faults appropriately.

That is successful troubleshooting.