On this page
Asset Type
Electrocardiograph (EKG) Machine
Manufacturer
Model
What This Guide Helps With
Troubleshooting a blank, dim, flickering, or distorted display caused by power, brightness, startup, external interference, connection, or display-system failure.
Step-by-Step Troubleshooting
1. Ensure Patient Safety and Stop Unreliable Use
Do not acquire or interpret an ECG when the display cannot be read reliably.
Disconnect the patient safely.
Transfer urgent testing to another verified electrocardiograph.
Do not navigate menus or enter patient information by memory.
Remove the unit from immediate clinical use if the display is intermittent.
Expected outcome: Patient identification, ECG acquisition, and device control do not depend on an unreadable display.
2. Confirm the Exact Display Condition
Determine whether the screen is:
- Completely blank
- Very dim
- Flickering
- Distorted or discolored
- Showing partial lines or missing areas
- Readable only at certain angles
- Normal during startup but faulty afterward
- Also note whether indicator lights, printer activity, sounds, or keypad responses suggest the device is otherwise operating.
Expected outcome: The issue is categorized as total loss, brightness problem, intermittent display, or image distortion.
3. Check AC Power and Device Operation
Confirm the power cord is fully connected.
Test the outlet using an approved method.
Observe power indicators.
Listen for normal startup activity.
Check whether the printer or other indicators respond.
Expected outcome: The device has stable power and the display problem is separated from a complete no-power condition.
4. Compare AC and Battery Operation
When safe:
- Start the unit on AC power.
- Observe the display.
- Test battery operation only if the battery is in serviceable condition.
- Note whether brightness or flicker changes between power sources.
- Stop testing if there is heat, odor, smoke, or battery damage.
Expected outcome: A power-mode-specific display problem is identified or ruled out.
5. Check User-Accessible Brightness or Contrast Controls
Use only normal operator-accessible controls.
Increase brightness gradually.
Verify the display is not set to an unusually low level.
Observe whether the setting remains stable.
Do not enter restricted service menus.
Expected outcome: The display becomes clearly readable and remains stable. If the setting was the cause, troubleshooting can stop after functional verification.
6. Restart the Electrocardiograph
When no ECG is being acquired, saved, or transmitted:
- Shut down normally when possible.
- Disconnect nonessential external accessories.
- Restart the device.
- Observe the display throughout startup.
Expected outcome: The display initializes normally after restart and remains readable.
7. Remove External Interference and Reposition the Unit
Move the unit away from:
- Strong electromagnetic sources
- Damaged power strips
- High-power equipment
- Direct sunlight or intense overhead glare
- Areas of extreme heat, cold, or moisture
- Confirm ventilation openings are unobstructed.
Expected outcome: Flicker or poor readability caused by the environment is eliminated.
8. Inspect the Display Housing and Controls
Inspect externally for:
- Cracks or impact damage
- Loose bezel or housing
- Liquid residue
- Pressure damage
- Discoloration
- Loose or sticking brightness controls
- Signs of overheating
- Do not press on the display panel or open the enclosure.
Expected outcome: The display assembly is externally intact. Visible physical or liquid damage requires removal from service.
9. Observe for Position-Related Intermittency
Without opening the device:
- Gently reposition the screen or device only as intended by its design.
- Observe whether flickering changes with normal movement.
- Do not twist, flex, or strike the housing.
Expected outcome: The image remains stable during normal positioning. Repeatable flicker suggests a service-level connection or display assembly problem.
10. Check Whether the Device Is Operating Without a Visible Display
If the screen is blank but the unit appears powered:
- Do not attempt blind patient operation.
- Note indicator lights, sounds, and printer response.
- Avoid repeated key presses that could alter data or configuration.
- Shut the unit down and remove it from service.
Expected outcome: A probable display-path failure is recognized without risking unintended operation.
11. Perform Final Functional Verification
After correction:
- Restart the unit several times.
- Confirm the display is bright, stable, and undistorted.
- Verify all screen areas are readable.
- Enter approved test patient data.
- Acquire and review an ECG simulator waveform.
- Confirm menus, saved records, and status messages display correctly.
- Complete required electrical-safety and functional testing after repair.
Expected outcome: The display remains fully readable during startup, acquisition, review, and record handling. The unit may be returned to service.
If the Problem Persists
Power, brightness, restart, accessory, environment, and external-damage causes have been ruled out. The remaining possibilities may include an internal display cable problem, backlight failure, display-panel fault, power-supply instability, graphics circuitry problem, or software issue.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate GE Healthcare documentation and approved test equipment
- Repaired or configured only by qualified personnel
- Do not continue using the device because the printer still works or because the image returns intermittently. Return the unit to service only after the display remains stable through complete functional testing.
- Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
A flickering or intermittently blank screen can cause incorrect patient entry or missed lead-quality warnings even when ECG printing still functions.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the MAC 5500 HD display was extremely dim and difficult to read during ECG acquisition."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found that the user-accessible display brightness had been reduced to its lowest practical setting."
Resolution
What action was taken.
Example:
"Clinical Engineering restored the brightness setting and verified a stable, readable display during startup, simulator acquisition, record review, and printing."
Helpful Details to Include (If Known)
- Blank, dim, flickering, or distorted condition
- Startup-screen behavior
- AC and battery comparison
- Brightness-control response
- Power indicator and printer activity
- Environmental conditions
- Housing, impact, or liquid damage
- Position-related intermittency
- Simulator and display-function results
- Final device disposition
Final Thought
Do not operate a diagnostic ECG system through an unreadable screen. Verify power, brightness, environment, and external condition first, then remove the unit from service when display reliability cannot be proven.
That is successful troubleshooting.