GE Healthcare MAC 7

Device Clock or Time Zone Does Not Synchronize With MUSE

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

GE Healthcare

Model

MAC 7

What This Guide Helps With

Troubleshoot incorrect MAC 7 date, time, or time-zone synchronization involving configuration, network access, synchronization sources, and MUSE communication.

Step-by-Step Troubleshooting

1. Protect Patient Record Integrity

An incorrect device clock can place inaccurate timestamps on ECG records and complicate patient-record matching.

If the displayed time is materially incorrect and accurate timestamps cannot be assured, use another verified ECG device until the issue is resolved or an approved clinical workaround is established.

Expected outcome: New ECG records are not knowingly created with unreliable timestamps.

2. Confirm the Exact Time Problem

Compare the MAC 7 clock with an approved facility time reference.

Determine whether the issue involves:

Document both the displayed device time and the reference time.

Expected outcome: The discrepancy is clearly defined. If the displayed time is actually correct, no corrective action is required.

3. Verify the Configured Time Zone

Using authorized configuration access, verify that the MAC 7 is assigned to the intended facility time zone and related approved time settings.

Do not alter time-zone settings solely to compensate for an unrelated synchronization problem.

Expected outcome: The configured time zone matches the facility standard. If correcting an authorized configuration error restores the correct displayed time, verify synchronization and stop troubleshooting.

4. Verify Network Connectivity

Because synchronization may depend on network services or MUSE communication, check the external network connection.

For wired systems:

For wireless systems, confirm normal connection status without modifying security credentials.

Expected outcome: The MAC 7 has stable network connectivity. If restoring connectivity also restores time synchronization, verify the clock and stop troubleshooting.

5. Verify MUSE Communication

Confirm whether other MUSE-dependent functions operate normally, such as approved order retrieval or ECG transmission.

If all MUSE communication is unavailable, treat the clock issue as potentially part of the larger communication failure rather than as an isolated clock problem.

Expected outcome: The scope of the issue is identified as time-only or part of a broader MUSE communication problem.

6. Compare With Another Networked Device

Check a known-good MAC 7 or other approved system using the same infrastructure, if available.

Determine whether other devices show the same time discrepancy.

A shared discrepancy may indicate an infrastructure, synchronization-source, or enterprise configuration problem rather than a MAC 7 hardware problem.

Expected outcome: Clinical Engineering determines whether the condition is isolated to one device or shared across the environment.

7. Verify Authorized Date/Time Configuration

Review the MAC 7 date/time and synchronization configuration against the approved facility standard or a known-good device.

Do not use undocumented service menus or manually override centrally managed settings without authorization.

Expected outcome: The device configuration is consistent with the expected synchronization method. If an approved correction restores synchronization, monitor through a normal restart and stop troubleshooting.

8. Perform a Restart and Persistence Check

When clinically appropriate, perform a normal shutdown and restart.

After startup:

1. Verify the displayed date and time. 2. Confirm the configured time zone. 3. Allow normal network/MUSE communication to establish. 4. Recheck the time against the approved reference. 5. Confirm that the correction persists.

Expected outcome: The MAC 7 maintains the correct date, time, and time zone after restart and synchronization. If so, troubleshooting can stop.

9. Verify ECG Timestamping

Using an approved test workflow and ECG simulator when appropriate, acquire a test ECG and confirm that the timestamp associated with the record reflects the correct date and time.

If MUSE transmission is part of the workflow, verify the receiving system reflects the expected timestamp.

Expected outcome: Device and downstream ECG timestamps agree with the approved time reference.

10. Escalate Persistent Synchronization Failure

If the MAC 7 repeatedly loses time, will not synchronize despite verified connectivity and configuration, or generates unreliable timestamps, remove it from service when accurate clinical record timing cannot be assured.

Expected outcome: Unreliable timestamping is prevented from affecting clinical documentation while the problem is escalated.

If the Problem Persists

If the time zone, authorized date/time settings, external network path, MUSE communication, and restart behavior are correct, common external causes have been ruled out.

Remaining categories may include device software, time-retention functions, enterprise time services, MUSE/interface configuration, network infrastructure, or another service-level problem.

The MAC 7 should be:

Coordinate with network, interface, or MUSE support if multiple systems exhibit the same discrepancy.

Before return to service, verify date, clock time, time zone, persistence after restart, ECG timestamping, and downstream record timing.

Knowing when to stop device-level troubleshooting and escalate a likely infrastructure or service-level synchronization problem is proper troubleshooting.

Clinical Use Tip

Verify the ECG timestamp as part of final testing because a correct-looking home screen clock does not by itself confirm correct patient-record timing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that ECGs from the MAC 7 displayed timestamps one hour different from the facility and MUSE time."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the MAC 7 configured for the wrong time zone while network and MUSE communication were functioning normally."

Resolution

What action was taken.

Example:
"Corrected the time-zone configuration using the approved facility setting, restarted the device, and verified correct device and transmitted ECG timestamps."

Helpful Details to Include (If Known)

Final Thought

Accurate timestamps are part of patient-record integrity; verify configuration and external synchronization before assuming an internal failure, then escalate persistent problems and document the results clearly.

That is successful troubleshooting.