GE Healthcare OEC Elite

DICOM Send, PACS, or Worklist Failure

On this page

Asset Type

C-Arm

Manufacturer

GE Healthcare

Model

OEC Elite

What This Guide Helps With

Troubleshoots image-send or worklist problems caused by network connectivity, destination availability, patient data, configuration, or infrastructure issues.

Step-by-Step Troubleshooting

1. Protect Clinical Workflow and Image Availability

Confirm images needed for immediate care remain available locally and notify clinical staff when PACS or worklist functionality is unavailable. Use the facility-approved alternate workflow for patient identification or image transfer when necessary.

Expected outcome: Patient care and image retention continue while the communication issue is investigated.

2. Define the Communication Failure

Determine whether the problem affects DICOM image send, modality worklist, both functions, or only one destination. Confirm whether all studies fail or only a particular patient or exam.

Expected outcome: The network problem is narrowed to a specific workflow.

3. Verify Local System Operation

Confirm the OEC Elite is otherwise functioning normally and images can be acquired, displayed, and retained locally as expected.

Expected outcome: The issue is confirmed as communication-related rather than a general workstation failure.

4. Check the Physical Network Connection

Inspect the network cable and accessible connection point for looseness, damage, broken latches, or disconnection. Verify the cable is connected to the intended network port.

Expected outcome: A valid physical network connection is present. If correcting the connection restores communication, proceed to final verification.

5. Check Link Status

Where externally visible, verify normal network link/activity indication at the device or network port. Compare with facility standards without altering switch configuration.

Expected outcome: The physical network path appears active.

6. Confirm the Scope of the Outage

Check with clinical staff or IT whether PACS, worklist, or other modalities are experiencing the same issue. Determine whether the destination service is currently available.

Expected outcome: The problem is identified as either local to the OEC Elite or part of a broader infrastructure outage.

7. Verify Patient and Exam Information

For failures affecting an individual study, confirm required patient and exam information is present and correctly entered. Do not alter completed clinical data without following facility procedures.

Expected outcome: The study contains the information needed for normal transmission.

8. Review Approved Network and DICOM Configuration

Compare operator-visible or approved configuration information with the facility's documented values. Do not change IP settings, DICOM destinations, application entities, ports, or service-level configuration without authorization.

Expected outcome: No obvious configuration discrepancy is identified, or an authorized correction restores communication.

9. Perform a Controlled Communication Test

Send an approved test study or repeat an authorized worklist query and verify receipt at the intended destination with the appropriate clinical or IT team.

Expected outcome: DICOM send or worklist functionality completes successfully. Troubleshooting can stop once the complete path is verified.

10. Escalate Persistent Network or DICOM Failure

If physical connectivity is normal but communication still fails, coordinate with PACS, networking, or vendor support as appropriate.

Expected outcome: The unresolved device-versus-infrastructure issue is escalated with useful test results rather than speculative configuration changes.

If the Problem Persists

Common local cable, connectivity, patient-data, destination-availability, and approved configuration causes have been ruled out. Remaining possibilities may involve network infrastructure, firewall or routing behavior, PACS services, DICOM configuration, application software, or device network hardware.

If local clinical operation is otherwise safe, disposition should follow facility policy for systems with unavailable connectivity. If loss of communication prevents safe workflow or required image retention, remove the OEC Elite from service and label it Out of Service until the issue is corrected.

Evaluate using appropriate GE Healthcare documentation and coordinate with PACS and network personnel. Only authorized personnel should change network or DICOM configuration. Verify successful end-to-end transfer before return to normal workflow. Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

A successful network link does not prove DICOM operation; verify that the intended study actually arrives at the correct PACS destination.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that studies from the OEC Elite remained local and were not reaching PACS."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the external network cable disconnected from the workstation network port."

Resolution

What action was taken.

Example:
"The network connection was restored and a test study was successfully transmitted and confirmed at PACS."

Helpful Details to Include (If Known)

Final Thought

Preserve images and clinical workflow first, then verify the complete path from physical network connection through patient data and destination availability before assuming a device failure. Coordinate infrastructure escalation appropriately and document end-to-end verification.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide