On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Use this guide for weak fluoroscopy, visibly irregular pulsing, or suspected output or dose-indication performance problems.
Step-by-Step Troubleshooting
1. Protect the Patient From Unverified Radiation Output
Stop clinical use if radiation output, pulsing, or dose indication appears unreliable. Do not troubleshoot radiation-performance problems while a patient is being exposed.
Expected outcome: The affected system is removed from patient imaging until output and indication performance can be verified.
2. Confirm the Reported Condition
Determine whether the complaint is poor image brightness, low apparent output, irregular pulse timing, abnormal dose indication, or a measurement obtained during physics or quality-control testing. Record the operating mode and displayed values without assuming the generator is defective.
Expected outcome: The exact performance concern is identified and reproducible under controlled conditions.
3. Inspect the Imaging Chain for External Causes
Verify that the detector is unobstructed, the C-arm is positioned correctly, the field is not unintentionally restricted, and no accessory, drape, table component, or dense object is unexpectedly within the beam path.
Expected outcome: No external attenuation or positioning problem explains the poor image. If correcting positioning restores expected image performance, troubleshooting can stop after verification.
4. Verify Imaging Mode and User-Accessible Settings
Confirm the intended fluoroscopy mode and other normal clinical selections. Compare them with a known normal setup without changing protected calibration or service parameters.
Expected outcome: The system is operating in the intended imaging mode. If an incorrect clinical selection caused the concern, correct it and verify normal performance.
5. Check Image Display Before Assuming Output Is Low
Inspect the display for brightness, contrast, source-selection, or monitor problems that could make a normally acquired image appear abnormally dark or unstable.
Expected outcome: Display performance is normal. If the perceived low output was actually a monitor issue, correct the display problem and separately verify imaging performance.
6. Observe Fluoroscopy Operation Under Controlled Test Conditions
Using an appropriate phantom or approved test object, initiate fluoroscopy and observe whether image brightness and pulse delivery appear stable. Do not use a patient as a test subject.
Expected outcome: Fluoroscopy appears consistent and repeatable. Obvious irregularity or unstable output requires continued investigation and likely removal from service.
7. Verify Output With Approved Radiation Test Equipment
If the complaint concerns radiation performance, use calibrated, appropriate diagnostic X-ray test equipment and the applicable manufacturer or facility procedure. Measure only parameters for which Clinical Engineering is qualified and equipped.
Expected outcome: Measured output and related parameters meet the applicable acceptance criteria. If they do, investigate display, workflow, or measurement setup rather than adjusting the generator.
8. Verify Dose-Indication Test Setup
If the dose or DAP reading is suspected, confirm the measuring instrument, geometry, detector placement, field size, and test setup are appropriate. Avoid comparing readings taken under different conditions.
Expected outcome: The independent measurement method is valid and repeatable. If the system indication agrees with the applicable verification method, the complaint is not reproduced.
9. Perform Repeatability Testing
Repeat the controlled acquisition or measurement enough times to determine whether pulsing, output, and indicated dose are stable rather than relying on a single exposure.
Expected outcome: Results are repeatable and acceptable. If repeatability is confirmed, troubleshooting can stop following documentation.
10. Escalate Unverified Radiation Performance
If output remains low, pulsing remains irregular, or dose indication cannot be verified, do not adjust internal generator calibration or radiation-control parameters without authorized service procedures.
Expected outcome: The system remains out of clinical service until evaluated by qualified personnel.
If the Problem Persists
External attenuation, positioning, mode selection, display behavior, and measurement setup have been ruled out. The remaining issue may involve generator regulation, tube performance, detector feedback, automatic brightness functions, dose-measurement circuitry, calibration, or another service-level imaging subsystem.
Remove the device from service, label it Out of Service, and send it for repair or bench evaluation. Use appropriate GE Healthcare documentation and calibrated approved test equipment. Radiation-output adjustment or calibration should be performed only by qualified personnel and in accordance with applicable facility and regulatory requirements.
Return-to-service testing should include appropriate radiation-performance, image-quality, and dose-indication verification.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Never use a patient to determine whether suspected fluoroscopy-output or pulse irregularity has been corrected.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that fluoroscopic images appeared unusually dark and the OEC One CFD seemed to be producing low output."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found that a dense table accessory had remained in the beam path during the reported examinations."
Resolution
What action was taken.
Example:
"Clinical Engineering repositioned the accessory and verified stable fluoroscopic imaging and repeatable output using an approved test setup."
Helpful Details to Include (If Known)
- Imaging mode in use
- Reported image appearance
- Pulse behavior observed
- Accessories or objects in the beam path
- Detector and C-arm positioning
- Test instrument used
- Test geometry
- Repeated measurement results
- Dose indication observed
- Final device status
Final Thought
Treat suspected radiation-output problems as performance and patient-safety issues, eliminate external causes first, verify with appropriate test equipment, and escalate rather than making unsupported generator adjustments.
That is successful troubleshooting.