Use this guide for image lag, geometric distortion, fixed lines, nonuniformity, or artifacts that may resemble a flat-field calibration problem.
Step-by-Step Troubleshooting
1. Protect the Patient From Unreliable Imaging
Do not continue using the system clinically if artifacts could obscure anatomy, mimic pathology, or affect procedural decisions. Move imaging to another verified system when necessary.
Expected outcome: Clinical decisions are not based on an unreliable image.
2. Characterize the Artifact
Determine whether the artifact is fixed or intermittent, moves with the C-arm, follows the detector, appears only during motion, occurs after high-contrast objects, or affects every image.
Expected outcome: The artifact pattern is clearly documented and reproducible.
3. Remove External Objects From the Beam Path
Inspect the table, detector area, covers, accessories, cables, drapes, protective devices, and nearby equipment for anything unintentionally entering the field.
Expected outcome: The beam path is clear. If the artifact disappears when the object is removed, verify several images and stop troubleshooting.
4. Clean Accessible Imaging Surfaces
Inspect and clean accessible detector covers and relevant external surfaces according to approved cleaning methods. Look for dried fluid, tape residue, markers, or contamination that could create fixed artifacts.
Expected outcome: External imaging surfaces are clean and undamaged. If the artifact is eliminated, verify repeated imaging.
5. Verify C-Arm and Detector Position
Place the system in a neutral, repeatable position and obtain an image using an approved uniform test object or phantom.
Expected outcome: The artifact can be assessed without unusual geometry or external attenuation.
6. Compare Different Imaging Conditions
Determine whether the artifact appears at multiple normal imaging modes and positions. Avoid changing protected detector calibration or service parameters.
Expected outcome: The issue is narrowed to a specific setup or is confirmed across normal imaging conditions.
7. Check for Display-Only Artifact
Review the same image on another available approved display or destination when practical. A display problem can mimic an acquisition artifact.
Expected outcome: The artifact is identified as originating in either the acquired image or the display path.
8. Perform Approved Detector Calibration Only if Authorized
If routine operator- or Clinical Engineering-accessible detector calibration is specifically authorized by the manufacturer and facility procedure, perform it exactly as documented. Do not enter restricted service calibration or invent a correction procedure.
Expected outcome: The calibration completes normally and a uniform test image no longer shows the artifact. If artifacts remain, do not repeatedly recalibrate.
9. Perform Final Image-Quality Verification
Use an approved test object or image-quality method to confirm uniformity, absence of fixed lines, acceptable lag behavior, and stable image presentation.
Expected outcome: Image quality is consistent and clinically usable. If successful, troubleshooting can stop.
10. Escalate Persistent Image Artifacts
If lines, lag, distortion, or nonuniformity persist after external causes and authorized calibration have been addressed, stop before internal detector repair or advanced calibration.
Expected outcome: The system is removed from clinical use for qualified imaging-system evaluation.
If the Problem Persists
External objects, contamination, positioning, display behavior, and any authorized routine calibration have been checked. The remaining issue may involve detector performance, calibration data, image-processing software, detector communication, generator synchronization, or another service-level imaging subsystem.
Remove the device from service, label it Out of Service, and send it for repair or bench evaluation. Evaluate using appropriate GE Healthcare documentation and approved imaging test equipment. Internal detector repair or protected calibration should be performed only by qualified personnel.
Before return to service, verify detector calibration as applicable, image uniformity, artifact-free imaging, and overall image quality.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
A reproducible detector line or fixed artifact should not be dismissed simply because the anatomy remains visible; it may still affect diagnostic or procedural interpretation.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported a persistent vertical line on OEC One CFD fluoroscopic images."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found dried contrast material on the accessible detector cover aligned with the reported artifact."
Resolution
What action was taken.
Example:
"Clinical Engineering cleaned the detector cover using the approved method and verified uniform, artifact-free images with an imaging test object."
Helpful Details to Include (If Known)
- Artifact type and location
- Whether it is fixed or intermittent
- Imaging modes affected
- Objects or accessories in the field
- Detector-cover condition
- C-arm positions tested
- Whether the artifact appears on exported images
- Calibration result if applicable
- Final image-quality verification
- Final device status
Final Thought
Characterize the artifact before assuming detector failure, eliminate beam-path and surface causes first, use calibration only when authorized, and verify image quality before return to service.
That is successful troubleshooting.