On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Use this guide when a touchscreen, hand control, footswitch, or related operator input does not respond consistently or at all.
Step-by-Step Troubleshooting
1. Protect the Patient From Unreliable Controls
Do not use an unreliable imaging, motion, or exposure control during a patient procedure. If the failed control affects fluoroscopy or C-arm movement, move the case to another verified system.
Expected outcome: No clinical procedure depends on an uncertain operator control.
2. Identify the Failed Control and Function
Determine whether the problem affects the touchscreen, a specific hand-control button, the complete hand control, one footswitch pedal, or the entire footswitch. Confirm which command is expected.
Expected outcome: The failed input and associated function are clearly identified.
3. Inspect the Control for Physical Obstruction or Contamination
Check the touchscreen and control surfaces for fluid, adhesive residue, drapes, covers, debris, or objects that could prevent normal operation. Inspect buttons and pedals for sticking or visible damage.
Expected outcome: The control surface is clean, unobstructed, and mechanically free. If cleaning or removing an obstruction restores normal operation, verify repeatability and stop troubleshooting.
4. Inspect External Cables and Connectors
Inspect the hand-control or footswitch cable, strain relief, connector, and accessible receptacle for damage, looseness, contamination, or bent contacts. Reseat normal external connectors.
Expected outcome: The accessory is securely connected. If function returns after reseating, repeat testing before release.
5. Compare With Another Normal Control Path
If the same function can safely be activated from another standard control, compare operation. This helps determine whether the problem is isolated to one accessory or affects the broader system.
Expected outcome: The fault is narrowed to the individual control or to the system.
6. Substitute a Known-Good Compatible Accessory
If an approved compatible hand control or footswitch is available, substitute it without changing system configuration.
Expected outcome: The known-good accessory operates normally. If it does, remove the defective accessory from service and verify the replacement before returning the C-arm to use.
7. Check Touchscreen Response Across Multiple Areas
For touchscreen complaints, test several normal user selections to determine whether the entire screen is unresponsive or only a particular region or software function is affected.
Expected outcome: Touch response is consistent across the normal user interface. Persistent dead zones or unintended selections require service evaluation.
8. Restart the System if the Failure Appears Software-Related
If the control is physically intact but the interface appears frozen, perform one normal controlled system restart.
Expected outcome: The interface returns to normal and remains responsive. If the freeze recurs, remove the system from service for further evaluation.
9. Perform Final Functional Verification
Verify every affected control through repeated operation under safe test conditions. For exposure controls, use an approved imaging test setup and confirm expected initiation and release behavior.
Expected outcome: Controls operate consistently without sticking, delay, unintended activation, or loss of response. If successful, troubleshooting can stop.
10. Escalate Persistent or Intermittent Control Failure
If an external control remains intermittent, fails with a known-good accessory, or produces unintended commands, stop troubleshooting before internal control electronics or protected configuration.
Expected outcome: The system or affected accessory is removed from service and referred for qualified repair.
If the Problem Persists
External contamination, accessory connections, individual controls, and basic software state have been checked. The remaining problem may involve internal input electronics, touchscreen hardware, communication circuits, application software, or service-level configuration.
Remove the affected equipment from service, label it Out of Service, and send it for repair or bench evaluation. Use appropriate GE Healthcare documentation and approved test equipment. Internal repair or protected configuration should be performed only by qualified personnel.
Return-to-service testing should verify all affected controls repeatedly and confirm that no unintended exposure or motion command occurs.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Intermittent exposure-control operation is unsafe even if the switch works most of the time; remove it from use until reliable operation is verified.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported that the OEC One CFD footswitch intermittently failed to initiate fluoroscopy."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found damage at the external footswitch cable strain relief and reproduced the failure when the cable was moved."
Resolution
What action was taken.
Example:
"Clinical Engineering replaced the footswitch with an approved compatible unit and verified repeated fluoroscopy activation and release using a controlled test setup."
Helpful Details to Include (If Known)
- Control affected
- Specific function affected
- Whether the problem is constant or intermittent
- Cable and strain-relief condition
- Connector condition
- Known-good accessory substitution
- Touchscreen areas tested
- Whether restart affected the problem
- Results of repeated control testing
- Final device status
Final Thought
Treat unreliable exposure and motion controls as safety-critical faults, check simple external causes first, use known-good substitutions when appropriate, and verify repeatable operation before release.
That is successful troubleshooting.