GE Healthcare Innova Series

Image Intensifier or Early Flat Panel Artifacts

On this page

Asset Type

Fluoroscopy / Interventional System

Manufacturer

GE Healthcare

Model

Innova Series

What This Guide Helps With

Identifies external causes of image artifacts such as lines, shading, distortion, or noise from connections, interference, or setup issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Identify Artifact Type and Pattern

Check Display Monitors First

Inspect Detector / Image Intensifier Connections

Inspect Detector Surface and Housing

Check for External Interference (EMI)

Verify System Grounding and Power Stability

Test Different Imaging Modes / Field Sizes

Check Collimator and Filter Positioning

Perform a System Reboot

Review Recent Changes

If the Problem Persists

If all external causes have been ruled out, the issue is likely internal to the detector, image intensifier, or processing electronics.

Remove the system from service, label Out of Service, and escalate for vendor service or advanced bench evaluation.

Knowing when to stop prevents unnecessary downtime and ensures safe escalation.

Clinical Use Tip

Do not troubleshoot imaging artifacts during an active case. Always move the patient to a properly functioning room before evaluating the system.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported vertical lines appearing on fluoroscopy images during procedures."

Cause

What was observed during troubleshooting.

Example:
"Loose detector signal cable identified causing intermittent artifact."

Resolution

What action was taken.

Example:
"Reseated and secured detector cable; verified normal image quality across all modes."

Helpful Details to Include (If Known)

Final Thought

Artifact troubleshooting should always begin with what is visible and accessible—connections, environment, and setup. This avoids unnecessary escalation and keeps systems available. When external causes are ruled out, prompt escalation protects patient safety and maintains workflow. Clear documentation ensures repeat issues can be quickly identified and addressed.

That is successful troubleshooting.