GE Healthcare Innova Series

Table Movement or Brake / Interlock Faults

On this page

Asset Type

Fluoroscopy / Interventional System

Manufacturer

GE Healthcare

Model

Innova Series

What This Guide Helps With

Identifies external causes of table movement failure, including brake engagement, interlocks, control issues, or obstructions affecting safe positioning.

Step-by-Step Troubleshooting

Ensure Patient and Staff Safety

Verify Table Power Status

Check Emergency Stop (E-Stop) Buttons

Inspect Brake Status and Controls

Test Table Movement from Multiple Controls

Inspect Interlocks and Safety Conditions

Check for Mechanical Obstructions

Inspect Cables and Connections

Power Cycle the System

Observe for Error Codes or Alarms

If the Problem Persists

If all external causes (power, controls, interlocks, obstructions) have been ruled out, the issue is likely internal (motor, brake assembly, control board, or position sensors).

Remove the system from service, label it Out of Service, and escalate for vendor service or qualified bench evaluation.

Recognizing when to stop prevents unsafe operation and further equipment damage.

Clinical Use Tip

Never troubleshoot table movement with a patient present. Always transfer the patient to a safe surface before evaluating movement or brake issues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Table would not move and brake appeared stuck during procedure setup."

Cause

What was observed during troubleshooting.

Example:
"Emergency stop button engaged, causing brake lock and interlock condition."

Resolution

What action was taken.

Example:
"Released E-stop, verified brake disengagement, and confirmed full table movement from all controls."

Helpful Details to Include (If Known)

Final Thought

Effective troubleshooting starts with safety and simple checks like E-stops, brakes, and interlocks. Many table issues are external and preventable. Knowing when to escalate is just as important as fixing the issue, and clear documentation supports long-term reliability and patient safety.

That is successful troubleshooting.