GE Healthcare MAC VU360

Printer Issue – Not Printing / Poor Print Quality

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

GE Healthcare

Model

MAC VU360

What This Guide Helps With

This guide provides a structured Clinical Engineering approach when a MAC VU360 does not print at all or produces faint, incomplete, or poor-quality ECG printouts.

The focus is on logical, external, and easily verifiable checks before assuming internal printer failure. These steps rule out common consumable and configuration issues while maintaining patient safety and minimizing unnecessary escalation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm Printer Is Enabled

Verify Paper Type and Installation

Inspect Paper Path and Compartment

Check Print Density / Contrast Settings

Perform Printer Self-Test

Check for System Error Messages

If the Problem Persists

If the printer still does not function after verifying settings, paper supply, density, cleaning, and performing a self-test, common external causes have been ruled out.

The issue is likely internal (thermal print head, feed motor, internal printer assembly, or connection issue).

Knowing when to stop prevents unnecessary downtime and protects equipment integrity.

Clinical Use Tip

Never troubleshoot printer components while a patient is actively being tested. If documentation is urgently needed, move the patient to another ECG machine and complete the study there. Troubleshooting should only occur when patient care is not impacted.

Work Order Documentation (CCR Method)

Complaint

What was reported by the clinical staff.

Example:
“ECG prints blank pages.”
or
“Printout is very faint and unreadable.”

Cause

What was observed during troubleshooting.

Example:
“Printer enabled and correct thermal paper loaded. Density adjusted with no improvement. Printer self-test prints blank output. Suspected thermal print head failure.”

Resolution

What action was taken.

Example:
“Removed unit from service and sent for repair.”
“Printer assembly replaced on bench. Self-test passed. Returned to service.”

Helpful Details to Include (If Known)

Final Thought

Printer issues are frequently caused by consumables, configuration, or simple mechanical obstructions. A methodical approach that starts with external checks prevents unnecessary escalation and supports accurate repair decisions. Clear documentation ensures continuity of care and proper follow-up.

That is successful troubleshooting.