GE Healthcare OEC 9900 Series

Error: COLL IRIS POTENTIOMETER ERROR - PRESS ANY KEY

On this page

Asset Type

C-Arm

Manufacturer

GE Healthcare

Model

OEC 9900 Series

What This Guide Helps With

This guide helps troubleshoot COLL IRIS POTENTIOMETER ERROR - PRESS ANY KEY on the GE Healthcare OEC 9900 Series C-Arm. This error indicates that the system is not receiving valid collimator iris position feedback. When this feedback is missing, incorrect, fixed, or unreliable, generator functions including X-rays may be disabled because the system cannot confirm proper collimator iris position.

Because collimation affects image field size and radiation exposure area, this error should be treated as a safety-related imaging control issue, not just a nuisance message.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check for Obvious External or Mechanical Problems

Check Collimator Iris Movement

Inspect Potentiometer Wiring and Connector Seating

Check Whether the COLL_IRIS_POSN Signal Changes

Isolate the Likely Feedback Path Cause

Clear the Error and Recheck Operation

If the Problem Persists

If the error continues after checking iris movement, visible mechanical condition, potentiometer wiring, connector seating, and COLL_IRIS_POSN signal behavior, external and common causes have been ruled out.

A persistent COLL IRIS POTENTIOMETER ERROR - PRESS ANY KEY may indicate a failed collimator iris potentiometer, wiring harness fault, Fluoro Functions PCB issue, or internal collimator control failure.

The device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot this error on an active patient. Move the patient to a backup device or approved alternate imaging method first. A C-Arm with unresolved collimator feedback problems may disable X-rays or create uncertainty about field size control.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"GE Healthcare OEC 9900 Series C-Arm displayed COLL IRIS POTENTIOMETER ERROR - PRESS ANY KEY. User reported X-rays or generator functions were disabled and collimator iris operation could not be confirmed."

Cause

What was observed during troubleshooting.

Example:
"Collimator iris position feedback fault observed. COLL_IRIS_POSN signal did not change normally with iris movement, indicating possible iris potentiometer, wiring, connector, mechanical iris, or Fluoro Functions PCB feedback issue."

Resolution

What action was taken.

Example:
"Removed C-Arm from service. Inspected collimator iris mechanics, potentiometer wiring, connector seating, and feedback signal path. Error persisted after external checks. Unit labeled Out of Service and escalated for repair or bench evaluation."

Helpful Details to Include (If Known)

Final Thought

The COLL IRIS POTENTIOMETER ERROR should be handled as a collimator position feedback failure. Start with patient safety, confirm the exact message, check external and mechanical causes first, inspect wiring and connector seating, then isolate whether the issue is potentiometer, harness, or board-related. If the feedback signal cannot be verified, removing the C-Arm from service is the correct decision.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide