GE Healthcare OEC 9900 Series

Error: COLLIMATOR STUCK

On this page

Asset Type

C-Arm

Manufacturer

GE Healthcare

Model

OEC 9900 Series

What This Guide Helps With

This guide helps troubleshoot the COLLIMATOR STUCK error on the GE Healthcare OEC 9900 Series C-Arm. This condition means the collimator iris cannot properly open or close, making iris adjustment unavailable or unreliable. Because collimation affects beam limitation, image quality, and patient dose control, the device should not remain in clinical use if the field size cannot be controlled safely.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check for Obvious External or Physical Causes

Check C-Arm Control Inputs

Power Cycle and Recheck

Inspect Accessible Cables and Connections

Check for Mechanical Binding Symptoms

Consider Iris Motor or Drive Failure

Consider Position-Feedback Problems

If the Problem Persists

If the COLLIMATOR STUCK error remains after checking for external obstruction, control input issues, restart behavior, visible cable problems, mechanical binding symptoms, and obvious damage, remove the C-Arm from service.

At that point, external causes have been ruled out and the issue may involve a failed iris motor, mechanical binding inside the collimator assembly, loose or damaged internal wiring, faulty collimator control circuitry, or position-feedback failure.

Clinical Use Tip

Never troubleshoot collimator movement on an active patient. Move the patient or procedure to a backup C-Arm first if beam limitation cannot be confirmed. A stuck collimator can affect radiation field control, image quality, dose management, and procedural workflow.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"User reported GE Healthcare OEC 9900 Series C-Arm displayed COLLIMATOR STUCK and the collimator iris would not open or close during adjustment."

Cause

What was observed during troubleshooting.

Example:
"Verified COLLIMATOR STUCK error during iris adjustment. No external obstruction or visible damage found. Error persisted after restart, indicating likely collimator iris motor, mechanical binding, wiring, control circuit, or position-feedback issue."

Resolution

What action was taken.

Example:
"Removed C-Arm from service and labeled Out of Service. Escalated for imaging service repair or bench evaluation of the collimator iris motor, wiring, control circuitry, and position-feedback components."

Helpful Details to Include (If Known)

Final Thought

The COLLIMATOR STUCK error should be treated as a beam-limiting and patient safety concern. Start with simple external checks, confirm the exact behavior, avoid forcing the collimator mechanism, and escalate when mechanical or internal control failure is likely. Clear documentation helps repair teams focus on the iris motor, mechanical movement, wiring, control circuitry, and feedback path.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide