GE Healthcare OEC 9900 Series

Error: DATA ERROR

On this page

Asset Type

C-Arm

Manufacturer

GE Healthcare

Model

OEC 9900 Series

What This Guide Helps With

This guide helps troubleshoot DATA ERROR on the GE Healthcare OEC 9900 Series C-Arm. This condition may occur when the system detects invalid data, corrupted stored information, ABS table problems, software mismatch, nonvolatile memory issues, or a data access failure. The system may shut down or fail to complete startup, preventing reliable fluoroscopy or imaging use.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Perform a Controlled Reboot

Check AC Power and Startup Conditions

Check the Workstation-to-C-Arm Connection

Check for Recent Service, Software, or Component Changes

Check for Repeated or Consistent Shutdown

Do Not Attempt Clinical Workarounds

If the Problem Persists

If DATA ERROR continues after reboot, stable AC power, cable checks, and review of recent service history, external and common causes have been ruled out.

A persistent DATA ERROR may indicate an internal software issue, stored-data problem, flash memory fault, nonvolatile memory corruption, ABS table problem, or workstation/C-Arm data compatibility issue.

Clinical Use Tip

Never troubleshoot this error on an active patient or during an active procedure. Move the patient or case to a backup imaging device first, then continue troubleshooting only after patient care is protected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"GE OEC 9900 Series C-Arm displayed DATA ERROR during startup and shut down. Staff unable to use C-Arm for imaging."

Cause

What was observed during troubleshooting.

Example:
"Confirmed DATA ERROR on boot. AC power and workstation/C-Arm interconnect cable checked and reseated. Error persisted, indicating likely corrupted nonvolatile data, software/data access issue, ABS table problem, or flash/workstation mismatch."

Resolution

What action was taken.

Example:
"Removed C-Arm from service and labeled Out of Service. Sent for bench evaluation/software-data troubleshooting. Software reload or data repair may be required before return to service."

Helpful Details to Include (If Known)

Final Thought

DATA ERROR on the GE Healthcare OEC 9900 Series should be treated as a serious system data fault, not just a nuisance startup message. Start with patient safety, confirm the error, check power and external connections, then stop once common causes are ruled out. Persistent data-related shutdowns require proper repair escalation and clear CCR documentation.

That is successful troubleshooting.