GE HEALTHCARE OEC 9900 SERIES

PACS / Network Communication Failure

On this page

Asset Type

C-Arm

Manufacturer

GE Healthcare

Model

OEC 9900 Series

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting a GE Healthcare OEC 9900 Series C-Arm that is unable to communicate with PACS or networked imaging systems. Symptoms may include:

The guide focuses on external and verifiable network issues first, before assuming internal device failure.

Step-by-Step Troubleshooting

Verify Network Connectivity

Check IP Configuration

Ping PACS / Server

Verify PACS Server Status

Check Firewall / Port Settings

Test DICOM AE Titles and Settings

Restart Network Equipment

If the Problem Persists

External network checks have been ruled out.

The device may have an internal network interface failure or corrupted network settings.

Knowing when to stop prevents unnecessary downtime and ensures patient safety.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“C-Arm unable to send images to PACS; error displayed on console.”

Cause

What was observed during troubleshooting.

Example:
“Device could not communicate with PACS server; network link and IP settings verified. Server reachable via other devices.”

Resolution

What action was taken.

Example:
“Device removed from service and labeled out of service. Escalated to GE service for network interface evaluation.”

Helpful Details to Include

Final Thought

Patient safety and logical troubleshooting are paramount. Start with simple network checks, document all observations, and escalate when internal issues are suspected. Proper CCR documentation ensures continuity and accountability.

That is successful troubleshooting.