GE Healthcare Optima IGS Series

Flat panel detector calibration or artifact issues

On this page

Asset Type

Fluoroscopy / Interventional System

Manufacturer

GE Healthcare

Model

Optima IGS Series

What This Guide Helps With

Identifies external causes of flat panel artifacts or calibration issues, including power instability, detector condition, and connection problems affecting image quality.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Identify Artifact Type and Behavior

Verify Artifact Across All Displays

Inspect Detector Surface and Housing

Check Detector Positioning and Mounting

Inspect Detector and System Cables

Verify System Power Stability

Check for Environmental Interference (EMI)

Perform Detector Calibration (If Accessible)

Restart the System

If the Problem Persists

If artifacts remain after completing these steps, common external causes have been ruled out. The issue is likely internal to the detector, acquisition electronics, or image processing system.

Remove the system from service, label it Out of Service, and arrange for vendor service or bench evaluation.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot detector or imaging issues on an active patient. Move the patient to another room or system before performing any image quality troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported vertical lines and shading artifacts appearing on all fluoroscopy images."

Cause

What was observed during troubleshooting.

Example:
"Detector calibration drift identified after ruling out cables, contamination, and display issues."

Resolution

What action was taken.

Example:
"Performed detector calibration; artifacts resolved and image quality restored."

Helpful Details to Include (If Known)

Final Thought

Effective troubleshooting starts with simple, external checks before assuming detector failure. Patient safety, structured logic, and clear documentation ensure reliable imaging systems and reduce unnecessary downtime or vendor calls.

That is successful troubleshooting.