GE Healthcare Panda Warmer

Error: Sys Fail 02 Heater Power Feedback Mismatch

On this page

Asset Type

Infant Warmer

Manufacturer

GE Healthcare

Model

Panda Warmer

What This Guide Helps With

Troubleshooting Sys Fail 02 when heater power feedback does not match commanded output, creating unreliable warming operation. Because the warmer cannot be trusted to deliver the intended thermal support while this error is active, the device must be evaluated away from patient use.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Inspect External Condition and Basic Setup

4. Confirm the Error Is Not Cleared by a Safe Power Restart

5. Run the Heat Engine Health Check in Service Mode

6. Check Accessible Connections Between the Control PCA and Power PCA

7. Check Connections Between the Power PCA and the Heat Engine

8. Determine Whether the Issue Has Been Resolved

If the Problem Persists

If Sys Fail 02 remains after external inspection, safe restart evaluation, accessible connection checks, and the Heat Engine Health Check, common externally correctable causes have been ruled out.

A persistent error indicates a failure in the heater power-control path, with the Power PCA identified as the likely repair focus based on the reported fault and manufacturer troubleshooting direction.

Do not return the warmer to clinical use with an unresolved heater power mismatch.

Clinical Use Tip

Never troubleshoot on an active patient. Move the infant to a backup warming device first, because an unresolved Sys Fail 02 means heater output cannot be trusted.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Panda Warmer displayed Sys Fail 02 during operation. Reported heater power did not match commanded heater power, making warming performance unreliable."

Cause

What was observed during troubleshooting.

Example:
"Heat Engine Health Check failed and Sys Fail 02 persisted after verifying the power source and inspecting accessible connections between the Control PCA, Power PCA, and heat engine. Findings indicate likely Power PCA failure."

Resolution

What action was taken.

Example:
"Removed warmer from clinical service, labeled unit Out of Service, documented failed heater power verification, and routed unit for bench repair and Power PCA evaluation/replacement."

Helpful Details to Include (If Known)

Final Thought

Sys Fail 02 should be treated as a heater reliability failure until proven otherwise. Protect the patient first, verify the reported condition, rule out accessible connection-related causes, perform the specified service check, and escalate appropriately when the error persists. Accurate CCR documentation supports safe repair decisions and a clear device history.

That is successful troubleshooting.