GE Healthcare Panda Warmer

Power Supply or Battery Issues

On this page

Asset Type

Infant Warmer

Manufacturer

GE Healthcare

Model

Panda Warmer

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting power supply or battery issues on the GE Panda Warmer. Symptoms may include the unit not powering on, intermittent shutdowns, alarms related to power or battery, or failure to charge the internal battery. The focus is on external checks before assuming internal failure.

Step-by-Step Troubleshooting

Verify AC Power Source

Check Power Cord and Connections

Inspect the Battery Indicator and Charge Status

Test with a Known Good AC Supply

Attempt Power-On Without Battery

Reset the Unit (if safe and supported)

If the Problem Persists

External checks (outlet, cord, connection, battery seating) have been ruled out. The power supply or internal battery is likely failing.

Knowing when to stop is proper troubleshooting—do not attempt internal board repair.

Clinical Use Tip

Do not attempt to troubleshoot or operate the warmer on a patient if the power supply or battery is unreliable. Transfer the infant to another verified warming device before performing troubleshooting. Ensure patient safety takes precedence over equipment troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Unit will not turn on; battery icon shows low despite being plugged in.”

Cause

What was observed during troubleshooting.

Example:
External AC power verified; power cord and connections intact; battery not charging.

Resolution

What action was taken.

Example:
Unit removed from service, labeled Out of Service, and sent for repair/bench evaluation.

Helpful Details to Include

Final Thought

Proper troubleshooting prioritizes patient safety and logical external checks first. Document each step clearly to ensure continuity and accountability, and escalate issues promptly when internal components fail. Removing the device from service when necessary prevents patient risk and protects staff.

That is successful troubleshooting.