GE Healthcare Vivid Series

DICOM/PACS Export Failure

On this page

Asset Type

Ultrasound System

Manufacturer

GE Healthcare

Model

Vivid Series

What This Guide Helps With

Troubleshooting failed image exports to PACS, DICOM send errors, or studies not appearing in the imaging system.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Prevents data loss and ensures patient care is not disrupted.
Why it matters: Prevents data loss and ensures patient care is not disrupted.

Verify the Reported Issue

Expected: Export fails or does not complete.
Why it matters: Confirms this is a reproducible DICOM communication issue.

Check Network Connectivity

Expected: Stable network connection.
Why it matters: Loss of network is the most common cause of DICOM failures.

Confirm DICOM Node Configuration

Expected: Correct and matching DICOM settings.
Why it matters: Even small mismatches prevent successful communication.

Verify DICOM Association (Echo Test)

Expected: Successful verification/echo response.
Why it matters: Confirms communication between ultrasound system and PACS server.

Check Export Queue / Storage Capacity

Expected: Queue processes normally and storage has space.
Why it matters: Full storage or stuck queues can block exports.

Validate Study Data and Patient Information

Expected: Clean, complete study data.
Why it matters: Invalid data fields can cause DICOM rejection by PACS.

Check PACS Availability

Expected: PACS system is operational.
Why it matters: Issue may be external to the ultrasound system.

Restart System and Retry Export

Expected: Export completes successfully after restart.
Why it matters: Clears temporary communication or software faults.

If the Problem Persists

If all external causes (network, configuration, PACS availability, and storage) have been ruled out, the issue is likely internal to the system software or communication module.

Remove the system from service if critical workflow is impacted. Label the device Out of Service and escalate for vendor support or advanced diagnostics.

Clinical Use Tip

Always ensure studies are successfully sent and verified in PACS before deleting them locally. Never troubleshoot DICOM issues during an active patient exam.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ultrasound system unable to send studies to PACS; exams not appearing in imaging system."

Cause

What was observed during troubleshooting.

Example:
"Incorrect DICOM AE Title configuration preventing successful association with PACS server."

Resolution

What action was taken.

Example:
"Corrected DICOM configuration settings and verified successful export using echo test and study transmission."

Helpful Details to Include (If Known)

Final Thought

DICOM export failures are most often caused by configuration or network issues, not internal hardware faults. A structured approach—starting with connectivity and settings—ensures safe, efficient resolution. Escalation is appropriate once external causes are eliminated, reinforcing the value of disciplined troubleshooting and clear documentation.

That is successful troubleshooting.