On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting helium gas loss, leak alarms, pneumatic drive faults, or balloon inflation problems on a Cardiosave IABP.
Step-by-Step Troubleshooting
Ensure Patient Safety First
A gas loss or pneumatic drive fault can affect balloon inflation and counterpulsation support.
Action:
- Do not troubleshoot casually while the patient depends on active IABP support.
- Notify clinical staff immediately if the pump is alarming during therapy.
- Have staff assess the patient, hemodynamic status, augmentation, and need for alternate support.
- If support is compromised, follow facility policy for switching to another IABP or escalating emergently.
- Keep the patient connected only if clinical staff determine it is safe.
Expected outcome: Patient support is maintained safely before equipment troubleshooting continues.
Why it matters: A pneumatic drive failure can reduce or stop balloon inflation, which may directly affect patient stability.
Verify the Exact Alarm or Fault Message
Confirm whether the issue is a true gas leak, helium loss, pneumatic drive issue, or setup-related condition.
Action:
- Read and document the exact alarm or fault message.
- Note whether the alarm occurs:
- At startup
- During helium fill
- During therapy
- After patient movement
- After a catheter or tubing connection change
- Check whether the balloon is still inflating and deflating.
- Ask staff whether augmentation dropped, timing changed, or alarms repeated after reset.
Expected outcome: The reported issue is confirmed and separated from unrelated ECG, pressure, timing, or trigger problems.
Why it matters: “Gas loss” complaints may come from the helium circuit, catheter tubing, connector setup, or an internal pneumatic fault.
Check the Helium Supply Status
The IABP requires a usable helium source for normal balloon drive operation.
Action:
- Confirm the helium tank or cylinder is installed correctly.
- Check the helium level or supply indicator.
- Make sure the tank is not empty or near empty.
- Verify the tank is seated and secured correctly.
- If available, compare against a known-good helium tank or supply per facility procedure.
Expected outcome: The pump has an adequate helium supply available.
If resolved: If replacing or reseating the helium source clears the fault and normal operation returns, document the helium supply issue and stop.
Why it matters: Low or depleted helium supply can appear as gas loss, fill failure, or pneumatic drive trouble.
Inspect the External Helium and Pneumatic Connections
Loose or damaged external connections are common causes of gas loss alarms.
Action:
- Inspect the external catheter tubing path from the IABP to the patient connection.
- Check for loose fittings, partially seated connectors, cracked tubing, kinks, or visible damage.
- Confirm all luer-style or pneumatic connectors are fully seated.
- Look for tubing pulled tight, trapped under equipment, or stressed from patient movement.
- Do not manipulate the catheter insertion site unless clinically appropriate staff are involved.
Expected outcome: External connections are secure, intact, and not kinked or damaged.
If resolved: If reseating an external connection clears the alarm and therapy remains stable, document the loose connection and stop.
Why it matters: The pump may detect gas loss if helium escapes through a loose connector or damaged tubing path.
Check for Obvious Catheter or Balloon-Related Signs
A catheter or balloon issue can trigger gas loss alarms and requires clinical escalation.
Action:
- Ask clinical staff whether blood is visible in the catheter tubing.
- Check whether the pump has displayed any blood detection, catheter leak, or balloon-related alarm.
- Ask whether the patient was recently repositioned or transported.
- Do not continue troubleshooting as a routine equipment issue if a balloon leak or catheter issue is suspected.
Expected outcome: Catheter-related risk is identified or ruled out by clinical staff.
If blood, catheter leak, or balloon rupture is suspected: Escalate immediately to the clinical team and remove the IABP from routine troubleshooting workflow.
Why it matters: Blood or suspected balloon compromise is a patient safety concern, not just an equipment alarm.
Verify the Balloon Catheter Setup and Size Selection
Incorrect setup information may affect pneumatic performance and alarm behavior.
Action:
- Confirm staff selected the correct balloon catheter size and setup configuration.
- Verify the catheter is compatible with the pump and facility setup.
- Confirm the correct connection method is being used for the catheter and extension tubing.
- Review whether the issue began after a catheter change, setup change, or transport.
Expected outcome: The pump setup matches the catheter and clinical configuration.
If resolved: If correcting the setup clears the fault and normal assist resumes, document the incorrect setup and stop.
Why it matters: Incorrect catheter or setup selections can affect drive volume, fill behavior, and leak detection.
Inspect for Tubing Occlusion or Flow Restriction
Restrictions in the pneumatic path can cause drive faults even when there is no obvious leak.
Action:
- Check for kinks, sharp bends, compression, or pinched tubing.
- Look near bed rails, transport handles, cable wraps, and under the mattress edge.
- Confirm tubing is routed with enough slack for patient movement.
- Ensure no clamps or covers are restricting the pneumatic line.
Expected outcome: The pneumatic path is open and unrestricted.
If resolved: If repositioning tubing clears the fault and operation remains stable, document the restriction and stop.
Why it matters: A blocked or restricted gas path can prevent proper balloon inflation and trigger pneumatic alarms.
Check the Pump’s External Condition
Physical damage or contamination can affect pneumatic and connection reliability.
Action:
- Inspect the pump housing, helium tank area, connector ports, and patient connection area.
- Look for damaged ports, bent fittings, missing seals, residue, liquid intrusion, or impact damage.
- Check for unusual sounds from the pump during attempted operation.
- Note any unusual heat, smell, or repeated startup behavior.
Expected outcome: No obvious external damage or unsafe condition is present.
If damage, odor, heat, or liquid intrusion is found: Remove the pump from service and send for repair evaluation.
Why it matters: External damage around pneumatic connectors or the helium system may indicate a leak path or internal damage.
Restart Only If Clinically Safe and Appropriate
A controlled restart may clear a temporary setup or detection issue, but it must not interrupt patient support.
Action:
- Only restart the pump if clinical staff confirm it is safe.
- Do not power cycle during active patient support unless directed by clinical protocol.
- If available, move support to another IABP before deeper troubleshooting.
- After restart, observe whether the same gas loss or pneumatic drive fault returns.
Expected outcome: The device either resumes normal operation or confirms a repeatable fault.
If resolved: If the fault clears and does not return during safe observation, document the alarm history and final status.
Why it matters: Repeat faults suggest a real leak, catheter issue, or internal pneumatic system problem.
Compare With a Known-Good Setup When Available
A known-good comparison helps separate pump issues from tubing, tank, or accessory issues.
Action:
- Use only approved accessories and facility-approved test setup methods.
- Compare with a known-good helium source if appropriate.
- Compare external tubing or connection accessories if available and safe.
- Do not perform invasive patient-side troubleshooting beyond facility policy.
Expected outcome: The fault follows either the external setup/accessory or remains with the pump.
If the fault follows the accessory: Replace the defective external accessory and document the finding.
If the fault remains with the pump: Remove the pump from service for repair evaluation.
Why it matters: Swapping safe external components can prevent unnecessary pump repair while avoiding internal disassembly.
If the Problem Persists
If helium supply, external connections, tubing restrictions, catheter setup, and obvious accessory issues have been ruled out, the issue is likely internal to the pneumatic drive, valve system, pressure sensing, helium control, or leak detection circuit.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair, bench evaluation, or vendor service
- Replaced clinically with another available IABP if patient support is needed
Knowing when to stop is proper troubleshooting. An IABP with a recurring gas loss or pneumatic drive fault should not be returned to patient use without proper evaluation and verification.
Clinical Use Tip
Do not perform extended troubleshooting on a patient who is dependent on IABP support. Stabilize the patient first, move support to another pump if needed, and troubleshoot only when it is safe to do so. Therapy continuity matters more than keeping the same pump in use.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Cardiosave Hybrid / Rescue IABP reported gas loss alarm and pneumatic drive fault during patient support."
Cause
What was observed during troubleshooting.
Example:
"Found helium tank low and external pneumatic connector partially seated at the catheter connection."
Resolution
What action was taken.
Example:
"Reseated connector, replaced helium tank, verified alarm cleared during safe observation, and returned unit to service after clinical confirmation."
Helpful Details to Include (If Known)
- Exact alarm or fault message
- Whether the issue occurred during startup, fill, transport, or therapy
- Helium tank level/status
- Helium tank replaced or reseated
- Tubing and pneumatic connections checked
- Catheter size/setup verified
- Any visible blood in tubing reported by clinical staff
- Any kinked, pinched, or damaged tubing found
- Whether the fault cleared after reseating connections
- Whether the fault returned after restart
- Unusual sounds, heat, smell, or physical damage
- Environmental factors during use or transport
- Indicator lights, alarms, or status messages observed
- Final device status: returned to service, removed from service, or sent for repair
Final Thought
Gas loss and pneumatic drive faults on an intra-aortic balloon pump require a patient-first response. Start with safe external checks such as helium supply, tubing, catheter setup, and visible connections, but do not force the device back into service if the fault repeats. Clear documentation helps show what was checked, what was ruled out, and why repair escalation was appropriate.
That is successful troubleshooting.