Getinge Cardiosave Hybrid / Rescue

Gas Loss, Leak, or Pneumatic Drive Fault

On this page

Asset Type

Intra-Aortic Balloon Pump

Manufacturer

Getinge

Model

Cardiosave Hybrid / Rescue

What This Guide Helps With

Troubleshooting helium gas loss, leak alarms, pneumatic drive faults, or balloon inflation problems on a Cardiosave IABP.

Step-by-Step Troubleshooting

Ensure Patient Safety First

A gas loss or pneumatic drive fault can affect balloon inflation and counterpulsation support.

Action:

Expected outcome: Patient support is maintained safely before equipment troubleshooting continues.

Why it matters: A pneumatic drive failure can reduce or stop balloon inflation, which may directly affect patient stability.

Verify the Exact Alarm or Fault Message

Confirm whether the issue is a true gas leak, helium loss, pneumatic drive issue, or setup-related condition.

Action:

Expected outcome: The reported issue is confirmed and separated from unrelated ECG, pressure, timing, or trigger problems.

Why it matters: “Gas loss” complaints may come from the helium circuit, catheter tubing, connector setup, or an internal pneumatic fault.

Check the Helium Supply Status

The IABP requires a usable helium source for normal balloon drive operation.

Action:

Expected outcome: The pump has an adequate helium supply available.

If resolved: If replacing or reseating the helium source clears the fault and normal operation returns, document the helium supply issue and stop.

Why it matters: Low or depleted helium supply can appear as gas loss, fill failure, or pneumatic drive trouble.

Inspect the External Helium and Pneumatic Connections

Loose or damaged external connections are common causes of gas loss alarms.

Action:

Expected outcome: External connections are secure, intact, and not kinked or damaged.

If resolved: If reseating an external connection clears the alarm and therapy remains stable, document the loose connection and stop.

Why it matters: The pump may detect gas loss if helium escapes through a loose connector or damaged tubing path.

Check for Obvious Catheter or Balloon-Related Signs

A catheter or balloon issue can trigger gas loss alarms and requires clinical escalation.

Action:

Expected outcome: Catheter-related risk is identified or ruled out by clinical staff.

If blood, catheter leak, or balloon rupture is suspected: Escalate immediately to the clinical team and remove the IABP from routine troubleshooting workflow.

Why it matters: Blood or suspected balloon compromise is a patient safety concern, not just an equipment alarm.

Verify the Balloon Catheter Setup and Size Selection

Incorrect setup information may affect pneumatic performance and alarm behavior.

Action:

Expected outcome: The pump setup matches the catheter and clinical configuration.

If resolved: If correcting the setup clears the fault and normal assist resumes, document the incorrect setup and stop.

Why it matters: Incorrect catheter or setup selections can affect drive volume, fill behavior, and leak detection.

Inspect for Tubing Occlusion or Flow Restriction

Restrictions in the pneumatic path can cause drive faults even when there is no obvious leak.

Action:

Expected outcome: The pneumatic path is open and unrestricted.

If resolved: If repositioning tubing clears the fault and operation remains stable, document the restriction and stop.

Why it matters: A blocked or restricted gas path can prevent proper balloon inflation and trigger pneumatic alarms.

Check the Pump’s External Condition

Physical damage or contamination can affect pneumatic and connection reliability.

Action:

Expected outcome: No obvious external damage or unsafe condition is present.

If damage, odor, heat, or liquid intrusion is found: Remove the pump from service and send for repair evaluation.

Why it matters: External damage around pneumatic connectors or the helium system may indicate a leak path or internal damage.

Restart Only If Clinically Safe and Appropriate

A controlled restart may clear a temporary setup or detection issue, but it must not interrupt patient support.

Action:

Expected outcome: The device either resumes normal operation or confirms a repeatable fault.

If resolved: If the fault clears and does not return during safe observation, document the alarm history and final status.

Why it matters: Repeat faults suggest a real leak, catheter issue, or internal pneumatic system problem.

Compare With a Known-Good Setup When Available

A known-good comparison helps separate pump issues from tubing, tank, or accessory issues.

Action:

Expected outcome: The fault follows either the external setup/accessory or remains with the pump.

If the fault follows the accessory: Replace the defective external accessory and document the finding.

If the fault remains with the pump: Remove the pump from service for repair evaluation.

Why it matters: Swapping safe external components can prevent unnecessary pump repair while avoiding internal disassembly.

If the Problem Persists

If helium supply, external connections, tubing restrictions, catheter setup, and obvious accessory issues have been ruled out, the issue is likely internal to the pneumatic drive, valve system, pressure sensing, helium control, or leak detection circuit.

The device should be:

Knowing when to stop is proper troubleshooting. An IABP with a recurring gas loss or pneumatic drive fault should not be returned to patient use without proper evaluation and verification.

Clinical Use Tip

Do not perform extended troubleshooting on a patient who is dependent on IABP support. Stabilize the patient first, move support to another pump if needed, and troubleshoot only when it is safe to do so. Therapy continuity matters more than keeping the same pump in use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Cardiosave Hybrid / Rescue IABP reported gas loss alarm and pneumatic drive fault during patient support."

Cause

What was observed during troubleshooting.

Example:
"Found helium tank low and external pneumatic connector partially seated at the catheter connection."

Resolution

What action was taken.

Example:
"Reseated connector, replaced helium tank, verified alarm cleared during safe observation, and returned unit to service after clinical confirmation."

Helpful Details to Include (If Known)

Final Thought

Gas loss and pneumatic drive faults on an intra-aortic balloon pump require a patient-first response. Start with safe external checks such as helium supply, tubing, catheter setup, and visible connections, but do not force the device back into service if the fault repeats. Clear documentation helps show what was checked, what was ruled out, and why repair escalation was appropriate.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide