On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting CS300 helium leak, low helium pressure, refill, autofill, or gas supply issues from external causes.
Step-by-Step Troubleshooting
Ensure Patient Safety First
Helium pressure and leak issues can affect balloon inflation and therapy support.
Action:
- Remove the CS300 from patient use if the alarm occurs during active therapy and cannot be immediately corrected.
- Have clinical staff move the patient to another IABP if therapy is still required.
- Do not continue patient support on a pump with repeated helium loss, failed autofill, or suspected gas leak.
Expected outcome: Patient support continues on a reliable device.
Why it matters: A helium leak or low drive pressure can reduce balloon inflation performance and compromise therapy effectiveness.
Confirm the Reported Alarm or Symptom
Action:
- Ask staff what message appeared and when it occurred.
- Confirm whether the issue was low helium pressure.
- Confirm whether the issue was a helium leak alarm.
- Confirm whether the issue was autofill failure.
- Confirm whether the issue was a repeated helium refill request.
- Confirm whether there was loss of augmentation.
- Confirm whether the balloon was not inflating normally.
- Confirm whether the alarm occurred during startup, standby, or active pumping.
Expected outcome: The exact symptom is identified before troubleshooting.
Why it matters: A true helium system problem, patient catheter leak, empty tank, loose connection, or setup issue may appear similar at first.
Check Whether the Device Is Connected to a Patient
Action:
- If connected to a patient, do not disconnect helium or catheter connections without clinical direction.
- Coordinate with nursing, perfusion, anesthesia, or the provider responsible for IABP therapy.
- If not connected to a patient, continue with external checks.
Expected outcome: Troubleshooting is performed only when safe.
Why it matters: IABP troubleshooting can directly affect circulatory support.
Check the Helium Tank Supply
Action:
- Verify the helium tank is installed correctly.
- Check that the tank valve is fully open if the system uses an external or serviceable cylinder connection.
- Confirm the tank is not empty or near empty.
- Replace the helium cylinder if low, empty, expired, damaged, or questionable.
Expected outcome: The CS300 has a usable helium source.
Why it matters: Low or empty helium supply is one of the simplest external causes of low pressure or autofill failure.
Inspect the Helium Tank and Regulator Area
Action:
- Look for obvious damage, loose fittings, missing seals, cross-threading, or signs that the tank is not seated correctly.
- Listen for audible hissing around the tank/regulator area.
- Do not use oil, lubricant, or tape on oxygen or gas fittings.
Expected outcome: No obvious gas leak is present at the supply connection.
Why it matters: A poor cylinder connection can cause pressure loss before helium reaches the pneumatic system.
Power Cycle Only If Clinically Safe
Action:
- If the unit is not supporting a patient, power the CS300 down and restart it.
- Allow the device to complete startup checks.
- Observe whether the helium pressure or autofill error returns.
Expected outcome: A temporary startup or fill sequence issue clears.
Why it matters: If the fault returns immediately after restart, the problem is more likely a persistent leak, supply issue, or internal pneumatic problem.
If this resolves the issue, document the restart and final status, then stop.
Check the Balloon Catheter Connection
Action:
- Inspect the catheter helium line connection at the pump.
- Confirm the connection is fully seated and secure.
- Check for cracked fittings, loose adapters, damaged tubing, kinks, or liquid contamination in the helium tubing.
Expected outcome: The balloon catheter helium connection is intact and properly connected.
Why it matters: A leak or poor connection at the catheter line can cause repeated helium loss or autofill failure.
Inspect the External Tubing Path
Action:
- Trace the helium tubing from the pump to the catheter connection.
- Look for pinched tubing, stretched tubing, crushed tubing, loose connections, damaged connectors, or tubing caught in bed rails.
Expected outcome: The external tubing path is clear and undamaged.
Why it matters: A restricted or leaking helium path can interfere with filling and balloon inflation.
Determine Whether the Problem Follows the Catheter Setup
Action:
- If clinically appropriate and directed by the care team, determine whether the alarm appears related to the patient catheter, tubing, or pump.
- Do not swap patient-connected components casually.
- If the pump alarms with a connected catheter but passes checks without that setup, suspect catheter-side leak or connection issue.
Expected outcome: The likely source is narrowed to the pump or the patient/catheter side.
Why it matters: Repeated helium loss may be caused by a balloon catheter leak, not only the pump.
Check for Moisture or Blood in the Helium Line
Action:
- Inspect visible helium tubing and connectors for moisture, blood, discoloration, or contamination.
- If blood or fluid is seen in the helium pathway, remove the pump from service and notify clinical staff immediately.
Expected outcome: No contamination is present in the helium drive line.
Why it matters: Blood or fluid in the helium line may indicate catheter failure or balloon membrane compromise and is not a simple equipment reset issue.
Try a Known-Good Helium Tank If Available
Action:
- Install a known-good compatible helium tank.
- Restart the device if it is not in patient use.
- Observe whether the low helium or autofill alarm clears.
Expected outcome: The alarm clears if the original tank or tank connection was the issue.
Why it matters: This helps separate a simple supply issue from a pump-side pneumatic fault.
If this resolves the issue, document the tank replacement and final operational check, then stop.
Check for Repeated Autofill Attempts
Action:
- Observe whether the CS300 repeatedly attempts to fill or reports autofill failure after the supply and external connections are verified.
- Do not continue repeated resets if the alarm returns.
Expected outcome: Repeated autofill failure is treated as a persistent fault.
Why it matters: Repeated refill attempts may point to a leak, pressure sensing issue, valve issue, internal pneumatic failure, or catheter-side gas loss.
Remove From Service if the Fault Repeats
Action:
- If helium pressure remains low, the leak alarm repeats, or autofill continues to fail after external checks, remove the CS300 from use.
- Label it Out of Service.
- Send it for bench evaluation or vendor/service repair.
Expected outcome: The device is not returned to clinical use until the helium system is verified.
Why it matters: Helium delivery is core to IABP therapy. A recurring gas fault should not be treated as a nuisance alarm.
If the Problem Persists
If the helium tank, tank seating, external tubing, catheter connection, and visible leak points have been checked and the problem remains, common external causes have been ruled out.
The issue is likely internal to the helium delivery, pneumatic, valve, pressure sensing, regulator, or autofill system.
The CS300 should be removed from service, labeled Out of Service, and sent for repair, bench evaluation, or qualified service support.
Knowing when to stop is proper troubleshooting. A recurring helium leak, low helium pressure condition, or autofill failure should not be cleared repeatedly and returned to patient use without verification.
Clinical Use Tip
Do not troubleshoot a helium leak or autofill failure on an active patient unless the clinical team determines it is safe. If therapy is needed, move the patient to another IABP first. Treat repeated helium loss as a therapy reliability issue, not just an equipment inconvenience.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Clinical staff reported the Getinge CS300 displayed a low helium pressure/autofill failure alarm during setup."
Cause
What was observed during troubleshooting.
Example:
"Helium tank and external catheter tubing connections were checked; the alarm repeated after installing a known-good helium tank."
Resolution
What action was taken.
Example:
"Device was removed from service, labeled Out of Service, and sent for bench evaluation of the helium/pneumatic system."
Helpful Details to Include (If Known)
- Helium tank level checked
- Tank replaced or reseated
- Tank/regulator area inspected for hissing
- Alarm message observed
- When alarm occurred: startup, standby, or pumping
- Catheter helium line checked
- External tubing inspected for kinks or damage
- Any moisture, blood, or contamination noted
- Whether issue followed patient catheter setup
- Whether autofill repeatedly failed
- Final device status
Final Thought
For the CS300, helium system problems should be approached carefully because they directly affect balloon inflation and therapy support. Start with the obvious external causes, including helium supply, tank seating, tubing, and catheter connections. If the problem repeats after those checks, remove the pump from service and document the work clearly using CCR.
That is successful troubleshooting.