Getinge ED-Flow

Cycle Record, Printer, or Network Export Is Missing

On this page

Asset Type

Endoscope Reprocessor (AER)

Manufacturer

Getinge

Model

ED-Flow

What This Guide Helps With

Troubleshoots missing cycle documentation caused by printer supplies, connections, destination settings, network communication, record storage, or export-path problems.

Step-by-Step Troubleshooting

1. Protect the Patient and Preserve Reprocessing Documentation

Do not assume a scope is cleared for clinical use when required cycle documentation or traceability records are unavailable.

Preserve any available on-screen or local cycle information and follow facility policy for documenting affected reprocessing cycles.

Expected outcome: Clinical use and traceability decisions are not based on an unverified missing record.

2. Confirm What Is Actually Missing

Determine whether the problem involves:

Record the affected cycle, scope ID, date/time, and operator information when available.

Expected outcome: The failure is narrowed to record creation, printing, or external export.

3. Verify the Cycle Actually Completed and Was Stored

Using normal authorized controls, determine whether the affected cycle appears in local cycle history.

If the record exists locally, do not treat the problem as a missing-cycle-data failure; focus on printing or export.

If the record does not exist locally, preserve all available details and escalate appropriately.

Expected outcome: Local record availability is confirmed.

4. Inspect the Printer and Consumables

If a connected or integrated printer is involved, inspect:

Do not dismantle the printer mechanism.

Expected outcome: The printer is powered, loaded, connected, and free of obvious external obstruction.

5. Test Whether the Printer Responds

Use an authorized normal print function or another available record to determine whether the printer can produce output.

If another record prints normally, the problem may be isolated to the original cycle record.

If nothing prints, continue troubleshooting the printer or communication path.

Expected outcome: Printer functionality is distinguished from a specific missing record.

6. Inspect External Communication Connections

For networked record export, inspect accessible:

Reseat loose external connections if permitted.

Do not alter network parameters without authorization.

Expected outcome: The physical communication path is intact.

7. Determine Whether the Problem Is Local or Department-Wide

Check whether:

A problem affecting multiple systems strongly suggests infrastructure or server-side involvement.

Expected outcome: The issue is isolated to the ED-Flow or identified as a broader system problem.

8. Verify Export Destination and Normal Configuration

Using authorized operator or technical configuration views, confirm that the intended printer or network export destination remains selected.

Compare with a known-good ED-Flow configuration if available.

Do not change undocumented network, security, or application settings.

Expected outcome: The unit is configured to send records to the intended destination.

9. Check Date and Time When Relevant

Verify that the displayed date and time are reasonable.

Incorrect timestamps can make successfully exported cycle records difficult to locate in downstream systems.

Do not make configuration changes outside authorized procedures.

Expected outcome: Records can be searched using an accurate cycle timestamp.

10. Attempt a Controlled Record and Export Test

After correcting any printer, cable, or configuration problem, run the appropriate approved test or use an authorized existing test record.

Verify:

Expected outcome: Complete documentation is generated and reaches the required destination. Troubleshooting can stop after successful verification.

11. Preserve Missing-Record Information

If a specific cycle record cannot be recovered, document every available detail before further service work, including scope ID, cycle time, operator report, and displayed completion status.

Do not fabricate or manually reconstruct an official automated cycle record unless an approved facility process specifically allows it.

Expected outcome: Available evidence is preserved without creating misleading documentation.

12. Remove From Service When Required Records Cannot Be Reliably Produced

If required cycle documentation, printing, or export remains unreliable and the facility depends on that function for traceability, remove the ED-Flow from service until corrected.

Expected outcome: The reprocessor is not used when required documentation cannot be reliably maintained.

If the Problem Persists

Common external causes such as paper supply, printer loading, accessible cabling, destination selection, network connectivity, timestamp confusion, and local-versus-network record location have been ruled out.

Remaining causes may involve internal record storage, printer electronics, application configuration, network infrastructure, interface services, database communication, or another service-level condition.

The device should be:

Coordinate with IT, cybersecurity, the tracking-system administrator, or Getinge when the failure is outside the reprocessor's external connections.

Verify local record creation and the complete required documentation path before returning the unit to service.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Verify the complete record path from the ED-Flow to the required traceability destination rather than assuming a locally completed cycle was successfully documented elsewhere.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterile Processing reported that ED-Flow cycle records were visible locally but were no longer appearing in the networked tracking system."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found a loose Ethernet connection at the reprocessor with no stable network link."

Resolution

What action was taken.

Example:
"Reseated and secured the network connection, verified link restoration, completed a controlled record-export test, and confirmed the cycle record reached the tracking system."

Helpful Details to Include (If Known)

Final Thought

Protect traceability as part of patient safety, determine whether the failure is record creation, printing, or export, verify external causes before assuming internal failure, escalate appropriately, and document the final end-to-end verification.

That is successful troubleshooting.