Getinge ED-Flow

Disinfectant Dosing or Chemical Errors

On this page

Asset Type

Endoscope Reprocessor (AER)

Manufacturer

Getinge

Model

ED-Flow

What This Guide Helps With

Chemical level alarms, dosing errors, or failed cycles due to disinfectant supply, connection, or sensor-related issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: No patient care is delayed or compromised
Why it matters: Incomplete disinfection creates infection risk

Verify the Reported Error and Behavior

Expected: Clear identification of when dosing fails
Why it matters: Helps determine if issue is supply, detection, or delivery

Check Disinfectant Level (Most Common Cause)

Expected: Adequate chemical available for dosing
Why it matters: Low or empty containers trigger dosing failures

Verify Chemical Container Installation

Expected: Secure, correct connection with no air gaps
Why it matters: Loose or incorrect connections prevent proper suction

Inspect Pickup Tubing and Lines

Expected: Clear, unobstructed tubing path
Why it matters: Flow restrictions prevent accurate dosing

Check for Air in the Chemical Line

Expected: Continuous fluid path without air gaps
Why it matters: Air prevents consistent pump operation

Verify Correct Chemical Type and Compatibility

Expected: Approved, in-date disinfectant
Why it matters: Incorrect chemicals may trigger system errors or improper dosing

Check Waste and Drain System

Expected: Proper drainage with no backup
Why it matters: Some systems halt dosing if drainage is impaired

Restart the System and Attempt a Test Cycle

Expected: Normal dosing and cycle progression
Why it matters: Confirms whether issue was temporary or persistent

If the Problem Persists

If all external causes have been ruled out, the issue is likely internal (e.g., dosing pump failure, internal sensor fault, or control issue).

Remove the unit from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop prevents unnecessary risk and repeated failed disinfection cycles.

Clinical Use Tip

Never rely on a questionable disinfection cycle. If a dosing error occurs, the scope must be reprocessed on a verified working AER before patient use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"AER displaying disinfectant dosing error and aborting cycles during processing."

Cause

What was observed during troubleshooting.

Example:
"Disinfectant container empty and pickup tubing partially disconnected, preventing proper chemical delivery."

Resolution

What action was taken.

Example:
"Replaced disinfectant container, resecured tubing connection, and verified proper dosing with successful test cycle."

Helpful Details to Include (If Known)

Final Thought

Effective troubleshooting starts with simple, external checks. Chemical issues are often supply or connection-related. Prioritize patient safety, verify proper disinfection, and escalate appropriately when internal faults are suspected.

That is successful troubleshooting.