Getinge ED-Flow

Drying Cycle Failure or Incomplete Drying

On this page

Asset Type

Endoscope Reprocessor (AER)

Manufacturer

Getinge

Model

ED-Flow

What This Guide Helps With

Failure of the drying cycle, wet scopes after processing, or drying phase aborts due to airflow, connection, or configuration issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient care continues safely using a fully processed scope
Why it matters: Moisture promotes microbial growth and compromises disinfection

Verify the Reported Issue and Cycle Behavior

Expected: Clear identification of when drying fails
Why it matters: Helps determine if this is a consistent failure or intermittent condition

Check Compressed Air Supply (Most Common Cause)

Expected: Stable air supply with audible airflow
Why it matters: Insufficient air pressure prevents proper channel and exterior drying

Inspect Air Filters and Dryer Components

Expected: Clean filters allowing unrestricted airflow
Why it matters: Reduced airflow leads to incomplete moisture removal

Verify Proper Scope Channel Connections

Expected: Secure, correct channel connections
Why it matters: Drying requires airflow through all internal channels

Check for Kinked or Blocked Tubing

Expected: Unobstructed tubing path
Why it matters: Blocked airflow prevents proper drying

Verify Drying Cycle Settings and Duration

Expected: Adequate drying time configured
Why it matters: Insufficient time may leave residual moisture even with proper airflow

Check Door Seal Integrity During Drying Phase

Expected: Tight seal with no air leakage
Why it matters: Air leaks reduce drying efficiency

Run a Test Cycle Without a Scope

Expected: Drying cycle completes without errors
Why it matters: Determines if issue is scope-related or system-related

If the Problem Persists

If airflow, connections, filters, and settings have been verified, the issue is likely internal (blower system, valves, sensors, or control failure).

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Do not attempt internal disassembly. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never assume a scope is safe if it appears only slightly damp. Even minimal retained moisture can compromise high-level disinfection. Always use a fully validated drying cycle.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported scopes coming out wet after ED-Flow cycle completion."

Cause

What was observed during troubleshooting.

Example:
"Clogged air intake filter restricting airflow during drying phase."

Resolution

What action was taken.

Example:
"Replaced air filter and verified proper airflow and complete drying cycle operation."

Helpful Details to Include (If Known)

Final Thought

Drying failures are most often airflow or connection related and can usually be resolved without invasive repair. Always prioritize patient safety by ensuring scopes are fully processed. When external causes are ruled out, escalation is the correct decision. Proper documentation supports both compliance and future troubleshooting efficiency.

That is successful troubleshooting.