What This Guide Helps With

Troubleshooting flow or pressure errors preventing proper channel flushing, often caused by blockages, connection issues, or supply restrictions.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: No patient risk and safe handling of contaminated equipment
Why it matters: Prevents exposure and ensures proper reprocessing workflow

Verify the Reported Error and Behavior

Expected: Clear understanding of when flow/pressure failure occurs
Why it matters: Helps isolate whether issue is supply, connection, or internal restriction

Check Scope Channel Connections (Most Common Cause)

Expected: Secure, correct connections with no looseness
Why it matters: Misconnections prevent proper flow through channels

Inspect for Kinked or Obstructed Tubing

Expected: Unrestricted, open tubing path
Why it matters: Even minor kinks can trigger pressure alarms

Check for Channel Blockage in the Scope

Expected: Free flow through all channels
Why it matters: Blocked channels will cause pressure buildup or flow failure

Verify Water Supply and Inlet Pressure

Expected: Stable and adequate water flow into the unit
Why it matters: Low inlet pressure prevents proper channel flushing

Inspect External Filters and Strainers

Expected: Clean filters with unrestricted flow
Why it matters: Partial blockage reduces system flow and increases pressure errors

Check for Air Traps or Improper Loading

Expected: Proper positioning with consistent flow
Why it matters: Air pockets can mimic pressure or flow issues

Restart the Cycle

Expected: Cycle runs without flow/pressure alarms
Why it matters: Confirms issue was resolved through external correction

If the Problem Persists

If all external causes (connections, tubing, filters, supply, and scope condition) have been ruled out, the issue is likely internal (pump, pressure sensor, or internal flow path restriction).

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop prevents unnecessary risk and wasted troubleshooting time.

Clinical Use Tip

Never troubleshoot flow or pressure issues with a contaminated scope actively connected. Always remove and reprocess safely before testing to avoid cross-contamination.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"AER reporting flow/pressure error during channel flush phase, cycle aborting."

Cause

What was observed during troubleshooting.

Example:
"Kinked scope channel tubing restricting flow and causing pressure alarm."

Resolution

What action was taken.

Example:
"Corrected tubing routing, verified proper connections, and confirmed normal cycle operation."

Helpful Details to Include (If Known)

Final Thought

Flow and pressure errors are most often caused by simple external restrictions or setup issues. A structured approach ensures patient safety, avoids unnecessary downtime, and supports efficient escalation when needed.

That is successful troubleshooting.