Getinge Flow-c

Device Will Not Power On or Complete Self-Test

On this page

Asset Type

Anesthesia Machine

Manufacturer

Getinge

Model

Flow-c

What This Guide Helps With

Troubleshoots startup, power, and self-test failures caused by external power, connections, accessories, configuration, or readily correctable setup conditions.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Anesthesia Capability

Do not troubleshoot an unreliable anesthesia machine while a patient depends on it. If the Flow-c is in clinical use, transition the patient to another verified anesthesia delivery and ventilation method according to facility procedures. Remove the machine from active use if dependable operation cannot be confirmed.

Expected outcome: Patient care is maintained independently of the affected machine, allowing safe troubleshooting.

2. Confirm the Exact Startup Failure

Determine whether the machine is completely unresponsive, begins powering up but stops, repeatedly restarts, or reaches a self-test that will not complete. Record any displayed message exactly as shown. Ask whether the problem began after relocation, cleaning, accessory replacement, shutdown, or a power interruption.

Expected outcome: The failure is reproduced and narrowed to power-up, initialization, or a specific self-test stage.

3. Verify Facility AC Power

Confirm the power cord is fully seated at the machine and connected to an appropriate receptacle. Inspect the plug and cord for damage. Verify the receptacle using an approved outlet tester or another appropriate method. Avoid relying only on the presence of nearby powered equipment.

Expected outcome: Reliable facility AC power is confirmed at the machine. If restoring AC power corrects startup, verify normal initialization and stop troubleshooting.

4. Inspect External Power Connections

Check the accessible power cord, strain relief, connector, and surrounding area for looseness, contamination, scorching, deformation, or physical damage. Do not use the machine with damaged power components.

Expected outcome: External power components are intact and securely connected. Damaged equipment is removed from service.

5. Check External Accessories and Connections

Inspect attached patient-circuit components, gas hoses, sensors, modules, data cables, and other removable accessories for improper connection or obvious damage. Correct any setup issue using approved components. Disconnect only accessories that can safely be removed and whose removal is permitted for troubleshooting.

Expected outcome: Required accessories are correctly installed and no external connection is preventing initialization.

6. Verify Required Gas and System Conditions

Confirm required pipeline or cylinder gas sources are available and properly connected as applicable. Check that accessible valves, hoses, and connections are in their normal operating condition. Do not bypass gas-supply safeguards.

Expected outcome: Required external gas supplies and connections are available for the machine's startup checks.

7. Perform One Controlled Restart

After external conditions have been corrected, shut down and restart the machine using the normal controls. Avoid repeated power cycling if the same fault continues.

Expected outcome: The Flow-c completes startup and its required self-test. If it does, continue to final verification and stop troubleshooting.

8. Perform Final Functional Verification

Before return to service, complete the applicable manufacturer-required pre-use or functional checks. Verify ventilation capability, gas delivery, alarms, controls, displays, and other safety functions appropriate to the machine.

Expected outcome: The machine completes required checks without unresolved faults and performs normally.

If the Problem Persists

If verified AC power, external connections, gas sources, accessories, and normal startup conditions have been ruled out, the problem may involve an internal power, battery, initialization, software, sensor, or other service-level condition.

Remove the Getinge Flow-c from service, label it Out of Service, and send it for repair or bench evaluation. Evaluate it using current manufacturer documentation and approved test equipment. Internal repairs, software recovery, or configuration changes should be performed only by qualified personnel.

Do not return the machine to clinical use until required functional and safety testing has been successfully completed. Recognizing when external troubleshooting has reached its safe limit is proper troubleshooting.

Clinical Use Tip

Have another verified anesthesia delivery and ventilation method available before troubleshooting any Flow-c that cannot reliably initialize.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"OR staff reported the Getinge Flow-c would power on but would not complete its startup self-test."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the AC connection partially disengaged at the machine, causing intermittent power during initialization."

Resolution

What action was taken.

Example:
"Clinical Engineering reseated the power connection, completed startup and functional checks, and verified normal operation before returning the machine to service."

Helpful Details to Include (If Known)

Final Thought

Protect the patient first, then verify facility power, external connections, gas supplies, accessories, and setup before assuming an internal failure. Escalate persistent startup faults appropriately and document the complaint, confirmed cause, corrective action, and final verification clearly.

That is successful troubleshooting.