Getinge Flow-e

Battery Will Not Charge or Support an Orderly Shutdown

On this page

Asset Type

Anesthesia Machine

Manufacturer

Getinge

Model

Flow-e

What This Guide Helps With

The internal battery does not charge, is not recognized, or cannot maintain the machine long enough for expected power-transition or shutdown behavior.

Step-by-Step Troubleshooting

1. Protect the Patient From Power Loss

Do not troubleshoot an anesthesia machine with unreliable backup power while it is supporting a patient.

If loss of AC power could interrupt ventilation, monitoring, or gas delivery, move the patient to a verified alternative system before testing the battery.

Expected outcome: Patient support is independent of the affected machine.

2. Confirm the Complaint

Determine whether the battery fails to charge, shows an abnormal status, immediately discharges during AC removal, or cannot support an orderly machine shutdown.

Expected outcome: The specific battery symptom is documented and reproducible.

3. Verify the AC Source

Confirm the Flow-e is connected to a known-good hospital-grade receptacle and that the receptacle has power.

Inspect the machine's power cord and plug for damage, looseness, overheating, or contamination.

Expected outcome: Reliable AC power is reaching the machine. If restoring AC power resolves the charge complaint, continue with battery verification before return to service.

4. Verify Charging Indication

With AC connected, observe the machine's normal battery and charging indications.

Allow charging under normal approved conditions rather than judging battery condition immediately after connection.

Expected outcome: The machine indicates that external power is present and the battery is charging or appropriately charged.

5. Inspect External Power Connections

Check accessible power-cord connections and strain reliefs for looseness or intermittent contact. Gently verify normal connection stability without stressing the cable.

Expected outcome: AC input remains stable when the cord is in its normal position. An intermittent or damaged cord requires replacement or repair before further testing.

6. Check Environmental Conditions

Confirm the machine has not recently been stored or operated in conditions that could adversely affect battery charging, such as unusual heat or cold.

Allow the device to return to an appropriate operating environment before retesting.

Expected outcome: Battery testing occurs under normal environmental conditions.

7. Perform a Controlled AC-to-Battery Transition Test

With the machine not connected to a patient and configured for approved bench testing, remove AC power and observe whether operation transitions appropriately to battery power.

Do not intentionally exhaust the battery outside approved service testing.

Expected outcome: The machine remains operational as intended during the transition and provides appropriate power-status indication.

8. Verify Orderly Shutdown Behavior

If the complaint specifically involves shutdown capability, verify according to the applicable manufacturer procedure that the battery supports the intended orderly shutdown sequence.

Expected outcome: The machine transitions and shuts down normally under the approved test procedure. If it cannot, remove it from service.

9. Inspect Battery Status Through Normal Service Information

Review battery condition information available through authorized normal technical procedures.

Do not modify protected battery parameters or use unauthorized service menus.

Expected outcome: Battery status is consistent with acceptable operation. A battery identified as degraded should be replaced using approved service procedures.

10. Complete Functional and Electrical Verification

After correction or approved battery replacement, verify charging, AC-to-battery transfer, battery operation, normal shutdown behavior, and applicable electrical-safety requirements.

Expected outcome: Power operation is reliable and all required tests pass before return to service.

If the Problem Persists

External AC source, power cord, operating environment, charging opportunity, and accessible power connections have been ruled out. Remaining causes may include a degraded battery, charging circuitry, internal power supply, battery interface, or other service-level power problem.

The device should be:

Do not rely on AC-only operation when the machine is expected to provide battery-supported transition or shutdown capability.

Return to service only after full power, battery, machine checkout, and applicable electrical-safety testing are successful.

Clinical Use Tip

Verify backup power before a case if the anesthesia machine may need to tolerate transfer, outlet interruption, or facility power disturbances.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Anesthesia staff reported that the Flow-e battery would not support the machine when AC power was disconnected during checkout."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering verified normal AC input and found the installed battery unable to maintain operation during controlled testing."

Resolution

What action was taken.

Example:
"Replaced the battery using the approved service procedure, verified charging and AC-to-battery transfer, completed checkout, and returned the machine to service."

Helpful Details to Include (If Known)

Final Thought

Start with the external AC source before blaming the battery, verify backup operation under controlled conditions, remove the machine from service when power transition is unreliable, and document both the correction and final power testing.

That is successful troubleshooting.