Getinge GSS Series Washer-Disinfector

Detergent Dosing Pump Failure

On this page

Asset Type

Washer-Disinfector

Manufacturer

Getinge

Model

GSS Series

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting detergent dosing pump failures on Getinge GSS Series washer-disinfectors. Failures may present as alarms indicating “dosing pump error,” lack of detergent delivery, or improper cleaning results. The guide focuses on external and easily verifiable checks before assuming internal pump failure.

Step-by-Step Troubleshooting

Verify Device Power and Status

Check Detergent Supply

Inspect Pump Connections and Electrical Supply

Run a Manual Pump Test (if available)

Check for Alarm History or Error Codes

Inspect Pump Mechanically

If the Problem Persists

Common external causes have been ruled out. The issue is likely internal to the dosing pump (motor, impeller, or internal electronics).

Reminder: Stopping troubleshooting at this stage is proper Clinical Engineering practice to ensure safety and avoid further damage.

Clinical Use Tip

Do not attempt to run cleaning cycles without detergent delivery; patient safety and infection control may be compromised. Move instruments to another operational washer-disinfector if possible before further testing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Detergent dosing pump failed; alarm displayed during cycle, no detergent delivered.”

Cause

What was observed during troubleshooting.

Example:
“Pump tubing and container were intact, electrical connections verified, pump does not actuate manually; internal pump failure suspected.”

Resolution

What action was taken.

Example:
“Washer-disinfector labeled Out of Service and sent for bench repair. Alternate washer used for patient instrument cleaning.”

Helpful Details to Include

Final Thought

Proper troubleshooting prioritizes safety, systematic logic, and knowing when to escalate. Documenting each step using CCR ensures clear communication and supports cost-effective maintenance while protecting patient care.

That is successful troubleshooting.