Getinge GSS67H

Chamber Will Not Fill, Drain, or Reach Level

On this page

Asset Type

Sterilizer

Manufacturer

Getinge

Model

GSS67H

What This Guide Helps With

Use this guide when chamber water or condensate handling is abnormal, drainage is incomplete, or expected process levels are not achieved.

Step-by-Step Troubleshooting

1. Protect Patient Care and Prevent Water Hazards

Do not process clinical loads when chamber filling or drainage is unreliable. If water is overflowing, leaking, or accumulating outside the sterilizer, stop operation and isolate the area as needed.

Expected outcome: Clinical loads are redirected and any immediate slip, electrical, or water-damage hazard is controlled.

2. Confirm the Exact Fluid-Handling Problem

Determine whether the chamber does not receive expected water, fills too slowly, fails to drain, retains condensate, or repeatedly stops during a fluid-related phase.

Document any displayed message and the phase where the problem occurs.

Expected outcome: The symptom is clearly identified rather than treating all fill and drain problems as the same failure.

3. Verify Facility Water Availability

Confirm required water supplies are available and that accessible facility valves associated with normal operation have not been unintentionally closed.

Coordinate with Facilities when the building supply is questionable.

Expected outcome: Required water service is available. If restoring an external supply corrects the problem, complete cycle verification and stop troubleshooting.

4. Inspect Accessible Supply Connections

Check external water hoses, piping interfaces, strainers accessible under approved maintenance procedures, and connections for kinks, leaks, obvious blockage, or damage.

Do not disassemble internal valves or piping assemblies.

Expected outcome: External supply paths are intact and unrestricted. If correcting an accessible restriction restores normal operation, verify and stop troubleshooting.

5. Inspect the Chamber Drain Area

When the chamber is cool and safe, inspect accessible drains and screens for debris, packaging material, labels, lint, or other obstruction.

Remove only material accessible without invasive disassembly.

Expected outcome: The drain path is visibly clear. If removal of debris restores drainage, perform an appropriate test cycle and stop troubleshooting.

6. Verify Facility Drainage

Check for evidence of backed-up building drains, slow drainage, nearby plumbing problems, or standing water at the sterilizer drain interface.

Coordinate suspected infrastructure issues with Facilities.

Expected outcome: Facility drainage accepts discharge normally. If a building drain condition is corrected and normal operation returns, verify a complete cycle and stop troubleshooting.

7. Check Loading Conditions

Confirm racks, containers, and load items are positioned normally and are not blocking drainage paths or causing excessive retained water.

Expected outcome: Load configuration is not interfering with chamber drainage or condensate movement.

8. Run an Appropriate Test Cycle

After supply and drain conditions are verified, run an approved test cycle and observe fill, conditioning, drain, and exhaust behavior.

Expected outcome: Chamber fluid handling progresses normally without standing water, repeated interruption, or abnormal leakage. Troubleshooting can stop after acceptable verification.

9. Escalate Persistent Fill or Drain Problems

If external supply, drain, chamber obstruction, and loading conditions are normal but the problem persists, remove the sterilizer from service.

Expected outcome: Internal fluid-control problems are referred for qualified service evaluation.

If the Problem Persists

Common external causes have been ruled out. Remaining categories may include internal valves, level sensing, pressure sensing, drain controls, piping, control outputs, or other service-level fluid-management components.

The sterilizer should be:

Stopping before unnecessary internal plumbing or component disassembly is proper troubleshooting.

Clinical Use Tip

Standing water or incomplete drainage can affect the sterilization process, so do not release loads from a cycle with unresolved chamber-fluid abnormalities.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterile Processing reported standing water remained in the Getinge GSS67H chamber after cycles."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found debris obstructing the accessible chamber drain screen."

Resolution

What action was taken.

Example:
"Clinical Engineering removed the debris, verified the drain path was clear, completed an approved test cycle, and confirmed normal chamber drainage."

Helpful Details to Include (If Known)

Final Thought

Start with safe isolation, facility water, drainage, accessible connections, and chamber obstructions before considering internal fluid-control faults. Verify normal operation with an appropriate cycle and document what changed.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide