Getinge GSS67H

Detergent, Chemistry, or Steam Delivery Fails

On this page

Asset Type

Sterilizer

Manufacturer

Getinge

Model

GSS67H

What This Guide Helps With

Use this guide when an installed process supply or steam source is unavailable, interrupted, or not delivered as expected during operation.

Step-by-Step Troubleshooting

1. Protect Patient Care and Process Integrity

Do not release a load from a cycle when a required process supply was unavailable or delivery was uncertain. Redirect processing to another validated device as needed.

Expected outcome: Affected loads are controlled and unreliable processing is prevented.

2. Confirm Which Supply Is Affected

Determine whether the reported issue involves steam or another configured process supply associated with the installed system.

Document the cycle phase and any displayed message.

Expected outcome: The failed supply path is accurately identified before troubleshooting begins.

3. Verify Supply Availability

For steam-related complaints, confirm facility steam is available. For any installed consumable or chemical supply, verify the correct approved product is present and not empty, disconnected, expired, or visibly damaged.

Expected outcome: The required process supply is actually available. If replacing or reconnecting an approved external consumable resolves the issue, continue to verification and stop troubleshooting.

4. Inspect Accessible Connections

Check accessible tubing, hoses, couplings, containers, and fittings for kinks, disconnection, leakage, incorrect seating, or visible damage.

Do not bypass sensing or delivery controls.

Expected outcome: External connections are secure and unobstructed. If correcting a connection restores delivery, verify operation and stop troubleshooting.

5. Verify Facility Steam Path

Inspect accessible external steam controls and coordinate with Facilities to ensure the steam path is open and stable.

Expected outcome: Steam reaches the sterilizer normally. If Facilities corrects the source condition and the sterilizer operates normally, complete cycle verification.

6. Check for External Leakage

Look for steam leakage, fluid leakage, condensation, or supply loss around accessible connections.

Avoid contact with hot piping and do not disturb pressurized fittings.

Expected outcome: No unsafe external leakage is present. Any significant steam or chemical leak requires removal from service.

7. Verify Correct Configuration Without Alteration

Confirm the intended cycle and installed supply configuration are being used. Do not change protected service parameters or substitute unapproved consumables.

Expected outcome: The device is being operated under its approved configuration.

8. Run an Appropriate Test Cycle

After external supply and connection issues are resolved, run an approved test cycle and verify that the required supply is delivered and the cycle completes normally.

Expected outcome: Supply delivery is consistent and the cycle completes without related alarms. Troubleshooting can stop after successful verification.

9. Escalate Unresolved Delivery Failure

If the supply is present, external connections are intact, and facility utilities are normal but delivery still fails, remove the sterilizer from service.

Expected outcome: Internal delivery-system faults are referred for qualified evaluation.

If the Problem Persists

Common external supply and connection problems have been ruled out. Remaining causes may involve internal valves, pumps where applicable, sensing, steam-control components, blocked internal pathways, or control-system faults.

The sterilizer should be:

Stopping before internal delivery-system disassembly is proper Clinical Engineering troubleshooting.

Clinical Use Tip

Never substitute an unapproved chemical, consumable, or alternate process configuration simply to make a sterilization cycle run.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterile Processing reported the Getinge GSS67H would not proceed normally because steam delivery was unavailable."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering confirmed the sterilizer was functional but found the external facility steam supply was unavailable."

Resolution

What action was taken.

Example:
"Facilities restored steam service, and Clinical Engineering completed an approved verification cycle with normal steam delivery and cycle completion."

Helpful Details to Include (If Known)

Final Thought

Identify the exact supply problem, verify external sources and connections first, and never bypass safety or substitute unapproved materials. Escalate internal delivery faults and document final process verification before return to service.

That is successful troubleshooting.