Getinge HS Series

Spray Arm Blockage

On this page

Asset Type

Washer-Disinfector

Manufacturer

Getinge

Model

HS Series

What This Guide Helps With

Identifies external causes of spray arm blockage that reduce cleaning effectiveness, such as debris, mineral deposits, or misalignment, preventing incomplete cycles.

Step-by-Step Troubleshooting

Ensure Patient and Staff Safety

Confirm washer is empty and staff are clear of potential spray.

Why: Prevents injury from sudden water release or chemical exposure.

Inspect Spray Arms for Obstructions

Remove upper and lower spray arms. Check for visible debris, detergent buildup, or mineral deposits in nozzles.

Expected outcome: Nozzles should be clear; debris can cause weak or uneven spray.

Clean Spray Arms

Rinse thoroughly under tap water; use a soft brush or toothpick to clear small holes.

Expected outcome: Water flows freely through all spray holes.

Check Spray Arm Rotation

Manually rotate the spray arms to ensure they move freely without obstruction.

Expected outcome: Spray arms rotate smoothly; restricted movement indicates blockage or mechanical interference.

Inspect Connections and Fittings

Verify spray arms are seated correctly on their mounts and that hoses or connectors are not kinked or loose.

Expected outcome: Proper alignment ensures consistent water distribution.

Run a Test Cycle

With washer empty, run a rinse cycle to observe spray arm rotation and water coverage.

Expected outcome: Spray arms rotate fully and no unusual noises are present; water reaches all areas of the chamber.

If the Problem Persists

All external causes have been ruled out; internal flow issues or pump problems may exist.

Remove the washer from service.

Label “Out of Service” and send for repair or bench evaluation.

Why: Continuing use could compromise cleaning efficacy and patient safety.

Clinical Use Tip

Never troubleshoot with instruments inside the washer.

If immediate cycle availability is required, transfer instruments to another washer-disinfector.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported poor cleaning in the upper tray and uneven spray."

Cause

What was observed during troubleshooting.

Example:
"Upper spray arm partially blocked with detergent residue; rotation restricted."

Resolution

What action was taken.

Example:
"Spray arm removed, cleaned, and reinstalled; test cycle completed successfully."

Helpful Details to Include

Final Thought

Thorough inspection and cleaning of spray arms ensures washer performance and patient safety. Proper documentation and knowing when to escalate protect staff and maintain device reliability.

That is successful troubleshooting.