Getinge Magnus Series

Table Drift or Inability to Hold Position

On this page

Asset Type

Surgical Table

Manufacturer

Getinge

Model

Magnus Series

What This Guide Helps With

Identifies causes of table drifting or failing to hold position, including hydraulic, lock, load, and control issues before assuming internal failure.

Step-by-Step Troubleshooting

Ensure Patient Safety

Verify Table Lock Status

Check Floor Conditions

Confirm Weight Limits and Distribution

Inspect for External Obstructions

Operate Table Through Full Range of Motion

Check for Audible Hydraulic Activity or Creep

Inspect for Visible Fluid Leaks

Cycle Power to Reset Control System

Test with Alternate Control (if available)

If the Problem Persists

If drift or inability to hold position continues after completing these steps, common external causes have been ruled out. The issue is likely internal, such as hydraulic valve failure or actuator issues.

Remove the table from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop is proper troubleshooting and protects patient safety.

Clinical Use Tip

Never troubleshoot drift while a patient is on the table. Even slight movement can compromise positioning, surgical accuracy, or patient safety. Always transfer first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Surgical staff reported table drifting downward during procedures."

Cause

What was observed during troubleshooting.

Example:
"No external issues found; suspected internal hydraulic pressure loss."

Resolution

What action was taken.

Example:
"Removed from service and sent for vendor repair."

Helpful Details to Include (If Known)

Final Thought

Table drift is a critical safety issue that requires a structured, external-first approach. Once load, environment, and controls are ruled out, escalation is appropriate. Proper documentation ensures clear communication and supports timely repair.

That is successful troubleshooting.