Getinge ROTAFLOW II

Display, Control Knob, or Alarm Audio Failure

On this page

Asset Type

Extracorporeal Membrane Oxygenation (ECMO) System

Manufacturer

Getinge

Model

ROTAFLOW II

What This Guide Helps With

Troubleshoots blank or abnormal display, unresponsive controls, and absent alarm audio involving power, connections, settings, damage, and environmental causes.

Step-by-Step Troubleshooting

1. Protect the Patient and Replace Unreliable User-Interface Equipment

A ROTAFLOW II with unreliable display, speed control, or alarm audio should not remain responsible for patient support when those functions cannot be trusted.

Coordinate transfer to verified backup equipment according to the ECMO team's contingency procedure.

Expected outcome: Patient support and alarm surveillance do not depend on an unreliable interface.

2. Confirm Which Function Has Failed

Determine whether the problem affects:

Record any intermittent behavior and displayed messages.

Expected outcome: The fault is isolated to a specific interface or alarm function.

3. Verify Console Power and Startup

Confirm the system has stable power and completes startup normally. A broader power problem can appear as a display or control failure.

Expected outcome: The console remains powered and stable. If power correction restores the interface, proceed to verification.

4. Inspect the External User Interface

Look for:

Do not operate equipment with a cracked enclosure, evidence of fluid intrusion, or damaged controls that could cause unpredictable operation.

Expected outcome: The interface is clean, dry, unobstructed, and physically intact.

5. Verify Normal User Settings

Check normal accessible settings relevant to display visibility or alarm presentation without changing protected clinical configuration.

Do not reduce, bypass, or disable required alarm functions in order to make the system appear operational.

Expected outcome: The complaint is not caused by an ordinary user-accessible setting.

6. Check Alarm Audio in an Approved Test

With the system safely off the patient, perform the normal alarm verification applicable to the equipment.

Confirm both audible and visual indications where required.

Expected outcome: Alarm audio is clearly present and corresponding visual indications operate normally. If audio remains absent, the device should not return to clinical use.

7. Exercise the Control Knob and Accessible Controls

On a safe test setup, operate the control knob and front-panel controls through normal user functions.

Look for:

Expected outcome: Each control produces a predictable single response.

8. Perform a Controlled Restart

If no physical damage is present and external causes have been excluded, perform one controlled restart.

Do not rely on repeated restarting to manage an intermittent user-interface or alarm failure.

Expected outcome: Display, controls, and alarm functions initialize and remain stable after startup.

9. Perform Final Functional Verification

Verify:

Expected outcome: User interface and alarm functions are reliable. Troubleshooting can stop after all return-to-service requirements pass.

10. Escalate an Unresolved Interface or Alarm Failure

If the display remains abnormal, controls remain unreliable, or alarm audio cannot be verified, remove the device from service.

Expected outcome: The unit is withheld from patient use until qualified repair restores reliable control and alarm functions.

If the Problem Persists

External power, settings, contamination, physical obstruction, and controlled restart have been ruled out. Persistent problems may involve the display assembly, controls, alarm hardware, internal communication, software, or system electronics.

The device should be:

Do not return an ECMO system to service with unreliable controls or alarm annunciation. Complete full functional and alarm verification following repair.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

A functioning pump is not enough for safe ECMO support; clinicians must also have reliable controls, displayed information, and audible alarm notification.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ECMO staff reported the ROTAFLOW II alarm could be seen on the display but could not be heard during equipment setup."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering confirmed alarm audio remained absent during approved testing despite normal power and user-accessible settings."

Resolution

What action was taken.

Example:
"Removed the console from service, labeled it Out of Service, and sent it for qualified repair because reliable alarm audio could not be verified."

Helpful Details to Include (If Known)

Final Thought

Protect the patient from unreliable controls or alarms, rule out power, settings, contamination, and external damage first, verify every corrected function, and remove the system from service when reliability remains uncertain.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide