Getinge ROTAFLOW II

Flow Sensor Not Detected, Will Not Zero, or Reading Is Incorrect

On this page

Asset Type

Extracorporeal Membrane Oxygenation (ECMO) System

Manufacturer

Getinge

Model

ROTAFLOW II

What This Guide Helps With

Troubleshoots missing, inaccurate, unstable, or non-zeroing flow measurements caused by sensor connection, placement, tubing, contamination, positioning, or setup.

Step-by-Step Troubleshooting

1. Protect the Patient and Confirm Flow Independently

Do not rely on a questionable flow value for patient management. Notify the ECMO clinical team and maintain support using approved alternate clinical information and backup equipment as required.

Do not manipulate a patient-connected ECMO circuit solely to troubleshoot a sensor.

Expected outcome: Patient management does not depend on an unreliable displayed flow measurement.

2. Confirm the Exact Flow-Sensor Problem

Determine whether the sensor:

Document the displayed value and operating conditions.

Expected outcome: The problem is specifically identified before components are moved or exchanged.

3. Inspect the Sensor and Cable Connection

Verify the flow sensor is securely connected and inspect accessible cable and connector surfaces for:

Do not continue using a damaged sensing cable.

Expected outcome: The connection is secure and the sensor is detected. If reseating the connection corrects the condition, proceed to verification.

4. Verify Sensor Placement

Confirm the flow sensor is installed on the appropriate circuit location and positioned according to the approved ROTAFLOW II setup.

Check for:

Expected outcome: The sensor is correctly positioned and securely coupled to the tubing.

5. Inspect the Tubing at the Measurement Site

Without disrupting patient support, or using a bench circuit when necessary, inspect the tubing where the sensor is applied for:

Expected outcome: The measurement site is appropriate and mechanically undistorted.

6. Verify Zeroing Conditions

If zeroing is required by the applicable operating procedure, ensure the system is placed in the proper safe condition before attempting the zero procedure.

Clinical personnel must manage any circuit condition required to create zero-flow conditions on patient-connected equipment.

Do not improvise clamps or alter ECMO flow solely for a technical test.

Expected outcome: The sensor zeros successfully when the correct clinical and manufacturer-defined conditions are present.

7. Test With a Known-Good Compatible Sensor

When available, test the console or flow-measurement channel using a known-good compatible sensor and suitable non-patient setup.

Avoid exchanging sensors on an active ECMO circuit unless directed by the clinical team and applicable procedure.

Expected outcome: Normal readings with a known-good sensor isolate the problem to the original sensor, cable, or its positioning.

8. Compare the Reading With the Test Setup

On a controlled bench setup, compare displayed flow behavior with expected changes in pump operation. Look for:

Do not invent calibration adjustments or enter restricted service functions.

Expected outcome: Flow measurement responds logically and remains stable.

9. Perform Final Functional Verification

After correction, verify:

Expected outcome: Flow monitoring is reliable and repeatable. Troubleshooting can stop after required return-to-service testing.

10. Escalate Persistent Flow Measurement Problems

If a known-good sensor, proper positioning, correct zeroing conditions, and secure connections do not restore reliable measurements, stop troubleshooting.

Expected outcome: The system is removed from service for flow-measurement channel or internal electronics evaluation.

If the Problem Persists

External sensor connection, placement, tubing condition, zeroing conditions, and known-good substitution have been ruled out. Remaining possibilities include a failed sensor channel, internal communication problem, configuration issue, or other service-level fault.

The device should be:

Following repair, complete manufacturer-appropriate flow monitoring and system functional verification before clinical release.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Treat an implausible ECMO flow display as unreliable until verified; never make circuit or patient-management decisions from a questionable single measurement.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ECMO staff reported the ROTAFLOW II flow sensor was detected but displayed an unstable flow value during setup."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the flow sensor incompletely seated around the tubing at the measurement location."

Resolution

What action was taken.

Example:
"Repositioned and fully seated the flow sensor, verified stable measurement response on the approved test setup, completed functional checks, and returned the system to service."

Helpful Details to Include (If Known)

Final Thought

Maintain safe patient support, verify sensor connection and placement before assuming a measurement-system failure, confirm readings under controlled conditions, and escalate unresolved inaccuracies.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide