Getinge ROTAFLOW II

Gas Blender or Oxygen Supply Problem in the ECMO Setup

On this page

Asset Type

Extracorporeal Membrane Oxygenation (ECMO) System

Manufacturer

Getinge

Model

ROTAFLOW II

What This Guide Helps With

Troubleshoots external ECMO gas-supply problems involving source gas, hoses, blender setup, oxygen connections, flow delivery, and associated accessories.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Gas Exchange

A loss or uncertainty of sweep gas or oxygen supply can immediately affect ECMO gas exchange. Notify the ECMO clinical team and establish a verified alternate gas source or approved backup method before technical troubleshooting.

Do not interrupt gas delivery to a patient-connected oxygenator for equipment testing.

Expected outcome: Required ECMO gas delivery continues through a verified clinical pathway.

2. Confirm the Exact Gas-Supply Problem

Determine whether the complaint involves:

Document the setup and any alarms or indicators.

Expected outcome: The affected portion of the gas-delivery chain is clearly identified.

3. Verify the Source Gas Is Available

Check the intended oxygen and other required source-gas connections used by the ECMO setup.

Verify:

Use approved hospital methods to verify medical gas infrastructure.

Expected outcome: Required source gas is present. If switching from a depleted or unavailable source to a verified source corrects the problem, proceed to final verification.

4. Inspect Gas Hoses and Connections

Trace the external gas path and inspect:

Do not use damaged medical gas hoses or improvised fittings.

Expected outcome: All hoses are correctly routed, intact, and securely connected.

5. Verify the Blender and Flow-Control Setup

Confirm the external gas blender, flowmeter, or related gas-control accessory is powered or supplied as applicable and that normal user controls are positioned appropriately.

Do not alter calibration or enter restricted service adjustments.

Expected outcome: The gas-control equipment receives the required source gases and provides an adjustable output.

6. Compare With a Known-Good Gas Source or Accessory

When safely possible, connect the setup to a verified alternate source or known-good compatible external gas-control component to determine whether the issue follows:

Coordinate all patient-connected changes with the ECMO clinical team.

Expected outcome: The failing portion of the external gas-delivery chain is isolated.

7. Inspect the Sweep-Gas Path to the Oxygenator

On a non-patient test setup, or clinically when managed by the ECMO team, confirm the outlet gas tubing is connected, unobstructed, and routed correctly to the oxygenator gas inlet.

Expected outcome: Gas can pass through the intended external delivery pathway without kinking, disconnection, or obstruction.

8. Verify Output With Appropriate Test Equipment

Where Clinical Engineering is responsible for the gas-control accessory, use approved test equipment and manufacturer procedures to verify output performance.

Do not make undocumented calibration adjustments based solely on a clinical reading.

Expected outcome: The external gas blender or flow-control device produces appropriate controllable output under the approved test procedure.

9. Perform Final Functional Verification

Confirm:

Expected outcome: The gas-delivery setup is stable and verified. Troubleshooting can stop.

10. Escalate an Unresolved Gas-Supply Problem

If the problem is traced to medical gas infrastructure, notify the appropriate facilities or medical-gas service team. If the blender or accessory fails verification, remove that component from service.

Expected outcome: An unreliable gas source or control device is not returned to patient use without qualified repair and verification.

If the Problem Persists

Common source-gas, hose, connection, blender setup, and sweep-path problems have been ruled out. Remaining causes may involve the medical gas infrastructure, gas-control accessory, regulator, configuration, or other service-level equipment.

The affected device or accessory should be:

Infrastructure problems should be escalated to the responsible hospital department. Complete appropriate gas-output and functional verification before returning the equipment to ECMO use.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Trace ECMO gas delivery from the source all the way to the oxygenator; a correct blender setting is meaningless if gas never reaches the circuit.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"ECMO staff reported no sweep-gas flow from the setup despite the external gas-control device being adjusted."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the oxygen supply hose incompletely connected at the wall medical-gas outlet."

Resolution

What action was taken.

Example:
"Secured the medical-gas connection, verified stable gas delivery and controllable output through the setup, and released the equipment after functional verification."

Helpful Details to Include (If Known)

Final Thought

Protect gas exchange first, trace the external gas path logically from source to oxygenator, verify equipment performance rather than guessing, and escalate infrastructure or service-level failures promptly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide