Getinge Servo-air Lite

High Airway Pressure Alarm

On this page

Asset Type

Ventilator

Manufacturer

Getinge

Model

Servo-air Lite

What This Guide Helps With

Addresses high-pressure alarms caused by circuit restriction, filters, water, expiratory obstruction, settings, accessories, or ventilator-related pressure-control problems.

Step-by-Step Troubleshooting

1. Protect the Patient First

Clinical staff should immediately assess the patient, airway, and ventilation when a high airway pressure alarm occurs. If effective ventilation is uncertain, move the patient to another verified ventilator or use an appropriate alternate ventilation method. Expected outcome: Patient ventilation is stabilized before equipment troubleshooting.

2. Reproduce the Condition on a Test Load

After removing the ventilator from patient use, connect an appropriate test lung and determine whether the high-pressure alarm recurs. Expected outcome: The alarm is either confirmed as an equipment/setup issue or does not reproduce outside the clinical condition.

3. Inspect the Circuit

Check inspiratory and expiratory tubing for kinks, trapped water, blocked connectors, pinched tubing, caps, or incorrect assembly. Expected outcome: The breathing circuit is unobstructed. If correcting an obstruction resolves the alarm, proceed to final verification.

4. Check Filters and Inline Accessories

Inspect filters, HMEs, humidification components, adapters, and other accessories for restriction or incorrect installation. Expected outcome: All components permit normal gas flow. Replace any restricted accessory.

5. Inspect the Expiratory Cassette

Confirm the cassette is fully seated, correctly prepared, and free from obvious external obstruction or damage. Expected outcome: The expiratory path is externally normal.

6. Review Settings

Review applicable pressure targets, pressure limits, PEEP, tidal volume, and alarm settings. Do not increase limits simply to eliminate an alarm. Expected outcome: Settings are appropriate for controlled bench testing and do not explain an expected high-pressure alarm.

7. Use a Known-Good Test Circuit

Replace the circuit and external accessories with known-good compatible components. Expected outcome: The alarm disappears if an external circuit component was responsible.

8. Evaluate Pressure Response

Observe pressure delivery with the test lung and use approved ventilator test equipment where required. Expected outcome: Pressure rises and stabilizes predictably. Uncommanded or excessive pressure requires removal from service.

9. Verify Alarm Performance

Create a controlled high-pressure condition to confirm the alarm activates and clears appropriately after the test condition is removed. Expected outcome: High-pressure alarm operation is normal. Troubleshooting can stop after successful testing.

10. Escalate Persistent High Pressure

If excessive pressure or the alarm persists with an unrestricted known-good circuit and appropriate settings, stop troubleshooting. Expected outcome: The ventilator remains out of service for qualified internal evaluation.

If the Problem Persists

Common external restrictions, filters, circuit components, cassette issues, and settings have been ruled out. The remaining cause may involve pneumatic control, pressure sensing, expiratory regulation, turbine control, calibration, or electronics. Remove the Servo-air Lite from service, label it Out of Service, and send it for bench evaluation using appropriate Getinge documentation and approved ventilator test equipment. Complete pressure, volume, alarm, and pre-use verification after repair. Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

A high-pressure alarm is first a patient and airway concern; equipment troubleshooting comes only after clinical causes and ventilation safety are addressed.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory Therapy reported recurrent high airway pressure alarms from the Servo-air Lite during setup."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found an occluded inline filter increasing breathing-circuit resistance."

Resolution

What action was taken.

Example:
"Clinical Engineering replaced the filter, verified normal pressure delivery and alarm operation with a test lung, completed the pre-use check, and returned the ventilator to service."

Helpful Details to Include (If Known)

Final Thought

Patient assessment comes first, followed by systematic external inspection and controlled testing before internal pressure-control failure is considered. That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide