Getinge Servo-i

Air or O₂ Gas Supply Failure

On this page

Asset Type

Ventilator

Manufacturer

Getinge

Model

Servo-i

What This Guide Helps With

Troubleshooting air or oxygen supply alarms caused by wall outlets, hoses, fittings, source pressure, crossed connections, or external gas-supply interruptions.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot a gas-supply failure while the Servo-i is actively supporting a patient.

Notify respiratory therapy and the clinical team immediately.
Transfer the patient to another verified ventilator or approved ventilation method.
Provide manual ventilation support when clinically required and directed by qualified staff.
Confirm adequate ventilation and oxygenation using independent monitoring.

Expected outcome: The patient is safely supported without depending on a ventilator with an unreliable gas supply.

Continue Clinical Engineering troubleshooting only after the Servo-i has been removed from patient use.

2. Confirm the Reported Gas-Supply Failure

Review the displayed alarm and determine whether the problem involves:

Record the exact alarm wording, when it occurred, and whether the alarm cleared after the device was disconnected from the patient.

Expected outcome: The affected gas source and failure pattern are clearly identified.

3. Inspect the Gas Hoses and Inlet Connections

Inspect the air and O₂ hoses from the ventilator to the source.

Check for:

Disconnect and firmly reconnect each hose to its correctly labeled ventilator inlet.

Expected outcome: Both hoses are secure, correctly routed, undamaged, and connected to the proper gas inlets.

If the alarm clears and the ventilator passes operational testing, stop troubleshooting.

4. Verify the Wall Gas Outlets

Confirm that each hose is fully seated in the correct wall outlet.

Test the suspected outlet using:

Do not assume that gas is available because the connector physically locks into the outlet.

Expected outcome: The wall outlet provides the correct gas at an acceptable and stable pressure.

If the outlet fails, remove it from clinical use and notify Facilities or the department responsible for the medical gas system.

5. Try a Known-Good Gas Outlet

Move the Servo-i to verified air and oxygen outlets in another location when practical.

Reconnect both gas supplies and restart the ventilator according to facility procedure.

Expected outcome: If the alarm clears at the alternate outlets, the original wall outlet or medical gas zone is the likely cause.

Stop troubleshooting the ventilator and report the facility gas-supply problem.

6. Substitute Known-Good Gas Hoses

Replace the air and O₂ hoses one at a time with compatible, approved, known-good hoses.

Do not use hoses with questionable fittings, improvised adapters, or uncertain gas identification.

Expected outcome: If the alarm clears after replacing a hose, the original hose or connector was defective.

Remove the failed hose from service and complete the required operational verification before returning the ventilator to use.

7. Check Portable Cylinder Supplies When Used

When the ventilator is connected to cylinders, verify:

A cylinder gauge may show remaining pressure while the regulator cannot maintain adequate flow under load.

Expected outcome: The portable gas source supplies stable pressure during ventilator operation.

8. Check for a Facility-Wide Gas Interruption

Determine whether other ventilators or gas-powered devices in the same area are reporting similar problems.

Contact Facilities or the medical gas system operator to check for:

Expected outcome: A facility-wide supply problem is either confirmed or ruled out.

9. Inspect the Ventilator Gas Inlet Area

With the ventilator disconnected from the patient, inspect the external gas inlet area for:

Do not insert tools into the gas inlets or attempt internal gas-module disassembly.

Expected outcome: The external inlet fittings are intact and show no evidence of leakage, contamination, or physical damage.

10. Restart and Observe the Alarm Sequence

With verified gas sources and known-good hoses connected:

A restart may clear a temporary detection condition, but it must not be used to dismiss a recurring supply alarm.

Expected outcome: Both gas supplies are detected normally and no supply alarm returns.

If the alarm clears, continue with a complete operational test before clinical use.

11. Perform an Operational Verification

Connect the ventilator to an approved test lung and verify:

Follow the facility’s approved inspection process and current manufacturer service documentation.

Expected outcome: The ventilator operates normally under test conditions without recurring gas-supply faults.

If the Problem Persists

If verified wall outlets, known-good hoses, correct connections, adequate source pressure, and external fittings have been confirmed, the common external causes have been ruled out.

The failure may involve an internal gas inlet assembly, pressure sensor, gas module, valve, wiring connection, or control system. Internal gas-system diagnosis requires appropriate service documentation, test equipment, training, and calibration procedures.

The ventilator should be:

Knowing when to stop and escalate is proper troubleshooting. Do not bypass gas-supply alarms or return the ventilator to service based only on an alarm temporarily clearing.

Clinical Use Tip

Never troubleshoot an air or O₂ supply failure while the Servo-i is supporting a patient. Transfer the patient first and independently verify ventilation and oxygenation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the Servo-i displayed an O₂ supply failure alarm and could not maintain the selected oxygen concentration."

Cause

What was observed during troubleshooting.

Example:
"The oxygen hose connector was worn and intermittently lost connection at the wall outlet; the ventilator operated normally with a known-good hose."

Resolution

What action was taken.

Example:
"Replaced the defective O₂ hose, connected the ventilator to verified air and oxygen sources, and completed test-lung and oxygen-concentration checks with no recurring alarms."

Helpful Details to Include (If Known)

Final Thought

Gas-supply alarms require immediate attention because they can affect ventilation and delivered oxygen concentration. Protect the patient first, verify external sources and hoses logically, and escalate when the problem points to the ventilator’s internal gas system. Clear CCR documentation preserves the evidence needed for safe repair and future trend review.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide