Getinge Servo-i

CO₂ Module Not Recognized or No Capnogram

On this page

Asset Type

Ventilator

Manufacturer

Getinge

Model

Servo-i

What This Guide Helps With

Troubleshooting a missing CO₂ waveform or unrecognized analyzer caused by module seating, sensor connections, airway-adapter contamination, warm-up, calibration, or configuration issues.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot unreliable CO₂ monitoring while the Servo-i is actively supporting a patient when capnography is clinically required.

Expected outcome: The patient is safely supported without relying on missing or unreliable Servo-i capnography.

Continue Clinical Engineering troubleshooting only after the device can be evaluated safely.

2. Confirm the Exact Failure

Determine whether:

Record all displayed messages before restarting or removing components.

Expected outcome: The failure is narrowed to module recognition, sensor detection, airway-adapter measurement, or calibration.

3. Verify That Only One CO₂ Module Is Installed

Inspect the Servo-i module compartment and confirm that only one CO₂ analyzer module is installed.

The Servo-i can handle only one CO₂ module at a time. Installing two modules can prevent normal recognition or operation.

Remove any duplicate module and restart the ventilator in accordance with facility procedures.

Expected outcome: One properly installed CO₂ module is present and recognized.

If recognition returns, complete an operational check and stop.

4. Reseat the CO₂ Module

Power the ventilator down after confirming it is removed from patient use.

Do not force the module or probe connector contacts.

Expected outcome: The Servo-i detects the installed CO₂ module during startup.

If the module remains unrecognized, test with a known-good compatible module when available.

5. Inspect the Capnostat Sensor and Cable

Check the external sensor assembly for:

Disconnect and reconnect the sensor cable securely.

Expected outcome: The capnostat sensor is fully connected and free of visible damage.

If a known-good compatible sensor restores operation, remove the failed sensor from service.

6. Allow the Sensor to Warm Up

After connecting the module and capnostat sensor, allow adequate warm-up time before evaluating accuracy or performing calibration.

The manufacturer notes that values displayed during sensor warm-up have reduced accuracy and that the sensor should be warm before calibration.

Expected outcome: The warm-up indication clears and the analyzer becomes ready for measurement.

If the waveform appears after warm-up, verify operation and stop.

7. Inspect the Airway Adapter

Remove the airway adapter from clinical use and inspect it under good lighting.

Check for:

If the Servo-i displays Check CO₂ airway adapter, confirm that the adapter is completely inserted and wipe or replace it as appropriate.

Use only an approved compatible airway adapter with the capnostat sensor.

Expected outcome: The adapter windows are clean, dry, unobstructed, and correctly seated.

If replacing the adapter restores the capnogram, document the contaminated or defective adapter and stop.

8. Verify Sensor and Adapter Orientation

Confirm that the capnostat sensor and airway-adapter windows are positioned vertically when connected to the test circuit.

Vertical placement helps reduce optical interference caused by contamination collecting on the adapter windows.

Expected outcome: The sensor is correctly positioned and produces a stable waveform during simulated ventilation.

9. Check for Nebulizer or Circuit Interference

Determine whether the reported failure occurred during nebulizer use or when excessive condensation was present.

Nebulizer operation can affect CO₂ readings, and moisture or deposits can obstruct the adapter’s optical path.

For bench testing:

Expected outcome: The waveform returns when optical interference and moisture are eliminated.

10. Verify CO₂ Display and Monitoring Configuration

Confirm that:

The Servo-i can display a CO₂ waveform, end-tidal CO₂, CO₂ minute elimination, and CO₂ tidal elimination when the analyzer is operating.

Expected outcome: The CO₂ parameter and waveform are visible in the configured display.

11. Perform the Approved CO₂ Verification or Calibration

Only perform calibration after the sensor has warmed and according to the approved Getinge procedure for the installed Servo-i software and analyzer version.

For systems using the documented CAPNOSTAT 5 procedure, the Servo-i provides a CO₂ calibration menu with a verification function using an unconnected airway adapter containing room air.

Do not perform improvised calibration or use unapproved calibration gases.

Expected outcome: The analyzer passes verification or calibration and produces a stable capnogram.

If calibration repeatedly fails, stop troubleshooting and escalate the device.

12. Substitute Known-Good External Components

When available, test one component at a time using compatible, verified parts:

Do not swap several components simultaneously because this prevents accurate identification of the cause.

Expected outcome: The defective external component is isolated.

After replacing the failed component, complete the manufacturer-approved functional and performance verification before returning the ventilator to service.

If the Problem Persists

If the Servo-i still does not recognize the CO₂ module, repeatedly loses communication, fails calibration, or produces no capnogram with known-good accessories, common external causes have been ruled out.

The problem may involve the CO₂ module, module connector, internal communication path, power supply, software configuration, or another internal assembly.

The ventilator should be:

Do not perform internal board-level repairs without appropriate training, documentation, test equipment, and authorization. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot the CO₂ module by repeatedly disconnecting the sensor or airway adapter while capnography is required for an active patient. Establish alternate verified ventilation and CO₂ monitoring first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the Servo-i did not display a CO₂ waveform and intermittently showed the CO₂ module as unavailable."

Cause

What was observed during troubleshooting.

Example:
"Found the capnostat airway adapter contaminated with condensation and not fully seated in the sensor."

Resolution

What action was taken.

Example:
"Replaced the airway adapter, reseated the capnostat sensor, allowed warm-up, and verified a stable capnogram and etCO₂ measurement during functional testing."

Helpful Details to Include (If Known)

Final Thought

CO₂ failures should be approached by protecting the patient first, then checking module seating, sensor connections, airway-adapter condition, warm-up, configuration, and calibration. Logical component substitution and accurate CCR documentation prevent unnecessary internal repair and support a safe return to clinical use.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide