Getinge Servo-i Ventilator

Fails Pre-Use Check / "Leak Detected"

On this page

Asset Type

Ventilator

Manufacturer

Getinge

Model

Servo-i

What This Guide Helps With

This guide addresses situations where a Getinge Servo-i ventilator fails its pre-use check and reports a “Leak Detected” error. The focus is on external and easily verifiable causes—such as breathing circuits, patient connectors, tubing, and accessories—before assuming internal hardware failure. The goal is to restore safe operation for patient use.

Step-by-Step Troubleshooting

Ensure Device Is Powered and Stable

Why: Low or unstable power can cause false leak detection.

Inspect the Breathing Circuit

Why: Even a small disconnection or crack can trigger a leak alarm.

Check the Patient Interface

Why: Leaks often occur at the interface rather than inside the ventilator.

Inspect the Exhalation Valve / Humidifier

Why: A misaligned valve or humidifier connection can mimic a leak.

Verify the Circuit Type Selected in Software

Why: Mismatched settings can cause pre-use check failures.

Perform a Manual Leak Test

Why: Confirms whether the leak is in the patient circuit or the ventilator.

Replace Consumable Components if Needed

Why: Consumables are common sources of leaks and are easily replaced without opening the ventilator.

If the Problem Persists

Common external causes have been ruled out. The ventilator may have an internal leak or valve malfunction.

Action:

Note: Knowing when to stop troubleshooting is proper Clinical Engineering practice.

Clinical Use Tip

Do not attempt on a patient: Always ensure the ventilator is removed from patient care during troubleshooting.

Use an alternative ventilator if available to maintain patient safety.

Troubleshoot only when safe and in accordance with hospital protocols.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pre-use check fails with ‘Leak Detected’ message.”

Cause

What was observed during troubleshooting.

Example:
Observed loose circuit connection at inspiratory limb. No cracks in tubing, valves seated correctly, no patient connected. Leak persisted after reconnecting.

Resolution

What action was taken.

Example:
Replaced disposable breathing circuit with a verified functional circuit. Pre-use check passed successfully. Ventilator returned to service.

Helpful Details to Include

Final Thought

Patient safety must always come first. Start troubleshooting from external, easily verifiable causes, escalate when necessary, and document every step clearly. Proper logic-based troubleshooting saves time, prevents unnecessary repairs, and ensures safe patient care.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide