Getinge Servo-i

O₂ Cell Calibration Failure or Incorrect FiO₂

On this page

Asset Type

Ventilator

Manufacturer

Getinge

Model

Servo-i

What This Guide Helps With

Troubleshooting failed O₂ cell calibration, oxygen-concentration alarms, or inaccurate FiO₂ caused by gas supply, connections, sensor condition, settings, or calibration problems.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot suspected oxygen-measurement or delivery problems while the Servo-i is actively supporting a patient.

Expected outcome: The patient is safely supported without relying on the affected ventilator.

Continue Clinical Engineering troubleshooting only after the Servo-i has been removed from patient use.

2. Confirm the Exact Failure

Review the reported symptom and determine whether the Servo-i:

Record the complete alarm wording and whether the problem is constant or intermittent.

Expected outcome: The failure is clearly identified as a calibration, measurement, or oxygen-delivery problem.

3. Inspect the Oxygen Supply

Verify that the oxygen hose is:

Confirm that the oxygen source is available and within the facility’s acceptable supply-pressure range. Test the outlet with an approved gas-pressure device or move the ventilator to a known-good oxygen outlet when appropriate.

Expected outcome: The ventilator receives a stable oxygen supply without leakage or restriction.

If the calibration succeeds after correcting the supply, document the finding and stop.

4. Verify the Air Supply

Confirm that the compressed-air hose is securely connected and that the air source is available and stable.

The Servo-i blends air and oxygen to produce the selected FiO₂. A missing, restricted, or incorrectly connected gas source may cause inaccurate oxygen delivery or prevent successful testing.

Expected outcome: Both gas sources are properly connected and available.

5. Allow the Ventilator to Stabilize

If the Servo-i was recently moved from a cold or hot environment, allow it to reach room temperature before repeating calibration.

Rapid temperature changes, condensation, or insufficient stabilization time may affect sensor performance.

Expected outcome: Environmental conditions are stable and suitable for calibration.

6. Inspect the O₂ Cell and Accessible Connections

With the ventilator powered off and removed from service, inspect the externally accessible O₂ cell installation according to facility-approved procedures.

Check for:

Do not open protected assemblies or proceed with internal disassembly unless authorized and trained under the manufacturer’s service program.

Expected outcome: The O₂ cell and its accessible connections are clean, dry, undamaged, and properly seated.

If reseating an accessible connection resolves the problem, repeat the complete pre-use check before returning the ventilator to service.

7. Check the O₂ Cell’s Age and Service History

Review the equipment history for:

Electrochemical oxygen cells are consumable components whose output decreases with age and use. A cell may become slow, unstable, or unable to reach the required calibration range before it fails completely.

Expected outcome: The O₂ cell is within the facility’s approved replacement interval and has no history suggesting deterioration.

8. Replace a Suspect O₂ Cell When Authorized

If external gas supplies and connections are verified but calibration continues to fail, replace the O₂ cell with the correct approved replacement part when permitted by facility policy and technician authorization.

Expected outcome: The replacement cell is recognized and produces stable oxygen measurements.

9. Run the Complete Pre-Use Check

Connect the Servo-i to known-good air and oxygen sources and perform the complete pre-use check, following all on-screen instructions.

Do not perform only a limited functional check. The automated pre-use process evaluates vital ventilator functions and provides calibration and status information before use.

Record:

Expected outcome: The Servo-i completes the pre-use check without an O₂ cell or gas-delivery failure.

If the complete check passes, continue with an independent oxygen-concentration verification before release.

10. Verify Delivered FiO₂ Independently

Connect the ventilator to an approved test lung and a calibrated external oxygen analyzer.

Test several oxygen settings appropriate to the facility’s performance-verification procedure, such as:

Allow readings to stabilize before comparing the Servo-i display, selected FiO₂, and independent analyzer measurement.

Inspect the test setup for leaks or entrained room air that could falsely lower the analyzer reading.

Expected outcome: Measured oxygen concentration remains stable and within the manufacturer’s applicable performance tolerance.

If the displayed or delivered FiO₂ remains inaccurate, remove the ventilator from service.

11. Evaluate Whether the Problem Follows the Cell

When facility policy permits, compare performance using a verified compatible O₂ cell or evaluate the suspect cell in another approved test setup.

Do not repeatedly substitute parts without documenting each test result.

Expected outcome: Testing distinguishes a failed consumable sensor from a ventilator-side fault.

12. Complete Final Safety Verification

Before returning the Servo-i to clinical service:

Expected outcome: Oxygen delivery, oxygen monitoring, alarms, and overall ventilator operation are verified as safe and accurate.

If the Problem Persists

If the O₂ cell calibration continues to fail or delivered FiO₂ remains inaccurate after verifying gas sources, hoses, connections, environmental conditions, sensor installation, sensor age, and an approved replacement cell, common external causes have been ruled out.

The problem may involve the internal oxygen-sensor interface, gas-blending system, pneumatic assembly, internal wiring, or control electronics.

The ventilator should be:

Do not return the Servo-i to patient use based only on a normal displayed value. Both the complete pre-use check and independent oxygen verification must pass. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot questionable oxygen delivery on an active patient. Transfer the patient first and verify oxygenation independently. A plausible FiO₂ display does not prove that the delivered oxygen concentration is accurate.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the Getinge Servo-i failed O₂ cell calibration and displayed FiO₂ values that did not match the selected oxygen setting."

Cause

What was observed during troubleshooting.

Example:
"Testing found an aged O₂ cell with unstable output; facility air and oxygen supplies, hoses, fittings, and accessible connections tested normally."

Resolution

What action was taken.

Example:
"Replaced the O₂ cell with an approved component, completed the full pre-use check, and verified delivered FiO₂ at multiple settings with a calibrated oxygen analyzer; unit passed and was returned to service."

Helpful Details to Include (If Known)

Final Thought

Accurate FiO₂ is essential to patient safety. Begin with gas supplies, hoses, connections, and sensor condition before suspecting internal failure. Complete pre-use testing, independent oxygen verification, appropriate escalation, and clear CCR documentation demonstrate sound Clinical Engineering practice.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide