Getinge Servo-i

Ventilator Will Not Start Ventilation After Standby

On this page

Asset Type

Ventilator

Manufacturer

Getinge

Model

Servo-i

What This Guide Helps With

Troubleshooting failure to resume ventilation caused by screen prompts, incomplete settings, circuit problems, gas supply, accessories, or system faults.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot a Servo-i that will not start ventilation while a patient is dependent on it.

Expected outcome: The patient is safely supported without relying on the affected ventilator.

Continue Clinical Engineering troubleshooting only after the Servo-i has been removed from patient use.

2. Confirm the Reported Behavior

Determine exactly what happens when the Start/Standby key is pressed.

Check whether:

Record the exact message, alarm text, and indicator-light behavior.

Expected outcome: The failure is clearly defined before components are changed.

3. Verify the Start/Standby Sequence

Press the Start/Standby key and follow the on-screen confirmation prompts. The Servo-i uses the Start/Standby control and a confirmation selection when changing between ventilation and Standby.

If ventilation begins normally, confirm stable operation with a test lung and stop troubleshooting.

Expected outcome: The ventilator leaves Standby and begins delivering breaths.

4. Check for Incomplete Patient or Ventilation Settings

Review the active patient category, ventilation mode, and required settings.

Confirm that:

Do not alter clinical settings from the reported configuration without coordinating with respiratory therapy. For bench testing, use facility-approved test settings.

Expected outcome: The ventilator has a complete and accepted configuration that permits ventilation to start.

5. Inspect the Patient Circuit

Connect a known-good, correctly assembled patient circuit and test lung.

Inspect for:

Reseat all accessible connections.

If ventilation starts with the known-good circuit, replace the defective circuit component and document the finding.

Expected outcome: The circuit is complete, unobstructed, and capable of supporting a controlled test breath.

6. Inspect and Reseat the Expiratory Cassette

Verify that the expiratory cassette is:

Remove and reinstall the cassette according to facility infection-control procedures. Do not use a cassette that is wet, cracked, contaminated, or physically damaged.

Expected outcome: The ventilator recognizes the expiratory cassette and permits ventilation.

7. Verify Air and Oxygen Supplies

Confirm that both gas hoses are fully connected to the ventilator and the facility gas outlets.

Check:

If available, verify source pressure using approved test equipment rather than relying only on the wall outlet appearance.

Expected outcome: Adequate air and oxygen supplies are available to the ventilator.

8. Check Electrical Power and Battery Status

Verify that the Servo-i is connected to a known-good hospital-grade outlet.

Confirm:

A ventilator may remain powered while still having a power-related condition that prevents reliable operation.

Expected outcome: The ventilator has stable AC power and no active power-system fault.

9. Review Active Alarms and System Messages

Open the alarm and message displays and record all active or recently generated conditions.

Pay particular attention to messages involving:

Do not clear or dismiss messages before recording them. The displayed information may identify the condition preventing ventilation.

Expected outcome: Any external condition blocking startup is identified and corrected.

10. Perform a Controlled Restart

With the ventilator disconnected from any patient:

Do not repeatedly cycle power if the unit displays a persistent technical error.

Expected outcome: A temporary software or control-state problem clears and the system starts normally.

11. Run the Pre-Use Check

Connect the required test equipment and run the Servo-i pre-use check according to facility procedure and the applicable manufacturer instructions.

Verify successful completion of:

Do not return the ventilator to service if the pre-use check fails, is interrupted, or cannot be completed.

Expected outcome: The ventilator passes its pre-use verification and starts ventilation with a test lung.

12. Test with Known-Good External Components

When compatible components are available, repeat the test using:

Change one component at a time so the actual cause can be identified.

If replacing an external component resolves the problem, repeat the pre-use check before returning the ventilator to service.

Expected outcome: The defective external component is isolated or external causes are ruled out.

If the Problem Persists

If the Servo-i still will not start ventilation after the controls, settings, circuit, expiratory cassette, gas sources, electrical supply, alarms, and pre-use check have been evaluated, common external causes have been ruled out.

The problem may involve an internal control, gas-delivery, sensor, power-supply, user-interface, or software fault.

The ventilator should be:

Do not perform internal board-level repair or bypass a startup interlock during routine floor troubleshooting. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a failure to start ventilation while the Servo-i is connected to an active patient. Move the patient to another verified ventilation method first and troubleshoot only under controlled conditions.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the Getinge Servo-i remained in Standby and would not begin ventilation when the Start/Standby control was selected."

Cause

What was observed during troubleshooting.

Example:
"Inspection found that the expiratory cassette was not fully seated, preventing the ventilator from completing the transition from Standby to active ventilation."

Resolution

What action was taken.

Example:
"Reseated the expiratory cassette, connected a test lung, completed the pre-use check successfully, verified stable ventilation and alarms, and returned the unit to service."

Helpful Details to Include (If Known)

Final Thought

A failure to leave Standby can result from a simple unconfirmed command, incomplete settings, an improperly installed circuit component, unavailable gas, or an internal fault. Protect the patient first, verify external conditions logically, complete functional testing, and escalate when the ventilator cannot pass its required checks. Clear CCR documentation preserves the troubleshooting history and supports safe repair decisions.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide