Getinge Servo-u Ventilator

Internal Battery Not Charging or Reduced Runtime

On this page

Asset Type

Ventilator

Manufacturer

Getinge

Model

Servo-u

What This Guide Helps With

This guide addresses situations where the Getinge Servo-u ventilator:

The focus is on logical, external power and environmental checks first before assuming internal battery failure.

Step-by-Step Troubleshooting

1. Ensure the Patient Is Safe First

2. Verify AC Power Source

Why this matters:
If the ventilator is not receiving consistent AC power, the battery will not charge.

3. Confirm Power Cord and Connection Integrity

A loose or damaged cord is a common and easily overlooked cause.

4. Check On-Screen Battery Status Indicators

If the display does not indicate charging while confirmed on AC power, further evaluation is needed.

5. Review Alarm History

Frequent power interruptions may explain reduced runtime due to incomplete charging cycles.

6. Inspect for Environmental Factors

Overheating can reduce battery life and charging efficiency.

7. Perform a Controlled Runtime Check (Bench Test Only)

Significantly reduced runtime suggests battery degradation.

8. Check for Recent Preventive Maintenance

Internal batteries have a limited service life and degrade over time.

If the Problem Persists

If:

Then common external causes have been ruled out.

The issue is likely:

Knowing when to stop prevents unnecessary risk and improper internal access.

Clinical Use Tip

Never troubleshoot internal battery concerns while the ventilator is supporting a patient.

If battery performance is questionable, transfer the patient to another ventilator before further testing.

Patient ventilation always takes priority over equipment evaluation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ventilator displayed low battery alarm and powered down during transport despite being plugged in earlier.”

Cause

What was observed during troubleshooting.

Example:
“Unit confirmed receiving AC power. Power cord and outlet verified functional. Battery indicated 100% but runtime test lasted only 8 minutes, indicating degraded internal battery.”

Resolution

What action was taken.

Example:
“Ventilator removed from service. Labeled Out of Service. Sent to manufacturer-authorized repair for battery replacement and charger system evaluation.”

Helpful Details to Include

Final Thought

Effective troubleshooting begins with simple external verification before assuming internal failure. Confirming power delivery, reviewing alarms, and performing controlled testing protects both the patient and the device. Recognizing battery end-of-life and escalating appropriately is responsible Clinical Engineering practice. Thorough CCR documentation ensures clarity and continuity of care.

That is successful troubleshooting.