Getinge Servo-u

Expiratory Cassette Test Fails After Cleaning or Replacement

On this page

Asset Type

Ventilator

Manufacturer

Getinge

Model

Servo-u

What This Guide Helps With

Helps isolate expiratory cassette test failures caused by incorrect installation, moisture, contamination, damage, cleaning-related issues, or cassette compatibility.

Step-by-Step Troubleshooting

1. Protect the Patient and Remove the Ventilator From Dependent Use

Do not troubleshoot a failed expiratory cassette test while a patient depends on the ventilator.

Provide alternate verified ventilatory support as clinically appropriate and remove the affected Servo-u from patient use until the expiratory system passes required testing.

Expected outcome: The patient is supported by a reliable device and the Servo-u can be evaluated safely.

2. Confirm the Exact Test Failure

Repeat only the normal user-accessible pre-use or system test appropriate to the expiratory cassette and note where the failure occurs.

Determine whether the problem began:

Do not repeatedly force the ventilator through failed testing if the same condition is consistently reproduced.

Expected outcome: The failure is confirmed and its relationship to cassette handling is known.

3. Verify the Cassette Is Fully Installed

Remove and reinstall the expiratory cassette using normal approved handling procedures.

Inspect that it:

Do not force the cassette into position.

Expected outcome: The cassette installs smoothly and is fully seated.

If correct installation allows the cassette test to pass, complete the remaining ventilator verification and stop troubleshooting.

4. Inspect for Residual Moisture

A cassette recently cleaned or processed may retain moisture that affects testing.

Inspect accessible areas for visible water droplets, condensation, or retained cleaning fluid. Allow the cassette to dry according to the approved cleaning and reprocessing process.

Do not use unapproved heat, compressed air, or drying methods.

Expected outcome: The cassette is visibly dry and ready for testing according to approved processing requirements.

If proper drying resolves the test failure, complete the full pre-use check and stop troubleshooting.

5. Inspect for Contamination or Cleaning Residue

Check accessible cassette surfaces and interfaces for:

Do not insert tools into protected sensing areas or attempt internal disassembly.

If contamination is present, have the cassette reprocessed according to the approved cleaning procedure.

Expected outcome: Accessible cassette surfaces and interfaces are clean and free of residue.

If proper reprocessing restores successful testing, complete functional verification and stop troubleshooting.

6. Inspect the Cassette for Physical Damage

Examine the cassette for:

A damaged expiratory cassette should not be returned to clinical service.

Expected outcome: The cassette is physically intact with no visible damage affecting fit or function.

If damage is found, remove the cassette from service and test the ventilator with a known-good compatible cassette.

7. Verify Cassette Compatibility

Confirm that the installed expiratory cassette is the appropriate compatible component for the Servo-u and has not been mixed with a cassette intended for another platform or configuration.

Avoid assuming similar appearance means compatibility.

Expected outcome: The cassette is confirmed to be appropriate for the Servo-u configuration being tested.

If an incorrect cassette was installed, replace it with the correct compatible cassette and repeat the required test.

8. Test With a Known-Good Compatible Cassette

Install a known-good compatible cassette that has been properly cleaned, processed, dried, and previously demonstrated normal operation.

Repeat the appropriate ventilator test.

Expected outcome: The ventilator successfully tests with the known-good cassette.

If the known-good cassette passes, remove the original cassette from service for further evaluation or replacement. Troubleshooting can stop after complete ventilator verification.

If the known-good cassette also fails, continue.

9. Inspect the Ventilator Cassette Interface

With the ventilator safely removed from patient use, inspect the accessible cassette receiving area for:

Do not disassemble the ventilator or manipulate internal sensors or valves.

Expected outcome: The cassette interface is clean, dry, unobstructed, and visibly intact.

If an external obstruction is safely removed and the cassette test passes afterward, complete the full pre-use verification and stop troubleshooting.

10. Perform Final Functional Verification

After resolving the problem, complete the required Servo-u pre-use and functional verification appropriate to the repair.

Confirm:

Expected outcome: The expiratory system performs normally and the ventilator is suitable for return to service.

11. Escalate Repeated Cassette Test Failures

If multiple known-good compatible cassettes fail despite correct installation and a clean, dry interface, stop external troubleshooting.

The ventilator should not be used clinically until the cause is evaluated by qualified service personnel.

Expected outcome: The device is properly removed from service for further evaluation.

If the Problem Persists

Common external causes such as installation, residual moisture, contamination, physical damage, processing issues, and cassette compatibility have been ruled out.

The remaining failure may involve the ventilator's expiratory cassette interface, sensing system, internal pneumatic components, protected calibration, or other service-level conditions.

The device should be:

Successful repair must be followed by the required pre-use testing and appropriate ventilator performance verification before return to patient care.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Never bypass a failed expiratory cassette test; move the patient to verified ventilatory support until the Servo-u passes all required pre-use checks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory Therapy reported that the Servo-u failed the expiratory cassette test after the cassette was cleaned and reinstalled."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found visible residual moisture in the recently processed expiratory cassette."

Resolution

What action was taken.

Example:
"Removed the wet cassette, allowed it to complete approved drying, reinstalled it, verified the cassette test and full pre-use check passed, and returned the ventilator to service."

Helpful Details to Include (If Known)

Final Thought

A failed cassette test should be approached from the outside in: verify processing, dryness, installation, compatibility, and the cassette itself before suspecting the ventilator. Protect the patient, verify the correction completely, escalate persistent failures, and document what was actually found.

That is successful troubleshooting.