Getinge Servo-u Ventilator

Expiratory Cassette Flow Sensor Failure or Calibration Errors

On this page

Asset Type

Ventilator

Manufacturer

Getinge

Model

Servo-u

What This Guide Helps With

This guide addresses situations where the Getinge Servo-u ventilator displays:

The focus is on external, logical checks first — cassette condition, moisture contamination, proper seating, and environmental factors — before assuming internal sensor or module failure.

Step-by-Step Troubleshooting

1. Remove the Patient from the Ventilator (If Applicable)

2. Verify the Exact Alarm or Error Message

3. Inspect the Expiratory Cassette

4. Check for Moisture or Condensation

5. Confirm Proper Cassette Seating

6. Run a New Pre-Use Check

7. Try a Known-Good Cassette

8. Inspect Cassette Interface Area

If the Problem Persists

If:

Then common external causes have been ruled out. The issue is likely internal to the ventilator’s measurement system or cassette interface.

At this point:

Knowing when to stop is proper troubleshooting. Avoid internal sensor or board-level disassembly.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ventilator failed pre-use check with “Expiratory cassette calibration failed” error. Unit unavailable for patient use.”

Cause

What was observed during troubleshooting.

Example:
“Moisture observed inside expiratory cassette. Calibration failed repeatedly until cassette replaced.”

Resolution

What action was taken.

Example:
“Replaced expiratory cassette with known-good unit. Pre-use check passed. Ventilator returned to service.”

Helpful Details to Include

Final Thought

Flow measurement errors on ventilators must be approached carefully and logically. Most issues involve cassette condition, moisture, or improper seating — not internal failure. Always rule out external causes first, protect the patient by removing the device from clinical use during troubleshooting, and escalate appropriately when internal faults are suspected. Clear documentation ensures continuity and regulatory compliance.

That is successful troubleshooting.