On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots supported Volista accessories or modules that are missing, unavailable, or intermittently recognized because of compatibility, connection, seating, or configuration issues.
Step-by-Step Troubleshooting
1. Protect the Patient and Identify the Affected Function
If the missing accessory supports a clinically required function, provide a verified alternative before troubleshooting. Do not disconnect or exchange equipment while it is actively required for patient care.
Expected outcome: Patient care continues safely while the accessory or module is evaluated.
2. Identify the Exact Accessory or Module
Record the accessory, module, interface, or optional component being reported as unavailable. Verify that it is intended for the specific Volista installation using labeling, asset history, or applicable manufacturer documentation.
Expected outcome: The affected component and expected system relationship are confirmed.
3. Confirm the Reported Recognition Failure
Determine whether the component is completely absent, intermittently detected, physically connected but unavailable from the normal interface, or became unrecognized after service, cleaning, relocation, or replacement.
Expected outcome: The symptom and its timing are clearly defined.
4. Inspect Physical Installation and Seating
Inspect the accessory and accessible mounting or connection points for incomplete seating, bent or damaged external contacts, contamination, strain, or obvious physical damage. Do not force connectors or dismantle internal modules.
Expected outcome: The component is correctly installed and externally undamaged. If proper seating restores normal recognition, verify function and stop troubleshooting.
5. Inspect Accessible Cables and Connectors
Check associated external communication or power cables for looseness, damage, pin problems, or incorrect routing. Reseat only approved user- or service-accessible connections with the system in a safe state.
Expected outcome: All accessible connections are secure and intact.
6. Verify Compatibility
Confirm that the accessory is approved for the Volista configuration and that no incompatible substitute has been installed. Compare model, part identification, or configuration information with applicable documentation without changing restricted system settings.
Expected outcome: The accessory is confirmed to be compatible with the installed system.
7. Use a Known-Good Comparison When Available
When safe and approved, test a known-good compatible accessory or test the suspect accessory on another compatible system. Do not exchange components in a manner that could alter critical configuration or calibration.
Expected outcome: The problem follows either the accessory or the host installation, narrowing the service path. If a defective external accessory is confirmed and replaced, verify operation and stop.
8. Perform a Normal Restart if Appropriate
After confirming proper installation, perform an approved normal restart if required for accessory detection. Do not access unauthorized service menus or force configuration changes.
Expected outcome: The supported accessory is recognized and remains available after restart.
9. Verify Complete Accessory Function
Recognition alone is not sufficient. Operate the accessory through all normal clinically used functions and verify stable communication and correct system response.
Expected outcome: The accessory is recognized consistently and functions correctly. Troubleshooting can stop.
10. Escalate an Unresolved Recognition Failure
If compatibility, seating, external connections, and known-good comparisons are satisfactory but the module remains unavailable, remove the affected function from service and escalate.
Expected outcome: The system is not returned to unrestricted use with an unresolved required accessory failure.
If the Problem Persists
Common external installation, compatibility, connection, and accessory causes have been ruled out. Remaining categories include installed wiring, system configuration, interface electronics, software, or an internal communication failure.
Remove the affected device or function from service, label it Out of Service when clinically required, and arrange qualified evaluation using current Getinge documentation and approved test equipment. Configuration changes and internal repair should be completed only by qualified personnel.
Before return to service, verify reliable detection and complete operation of the affected accessory. Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
After replacing or reconnecting an accessory, verify the actual clinical function rather than assuming that recognition alone proves proper operation.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported that a supported Volista accessory was no longer available from the system controls."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the accessory's accessible connector partially disengaged following equipment cleaning."
Resolution
What action was taken.
Example:
"Clinical Engineering reseated the connector, verified consistent accessory recognition and normal operation, and returned the system to service."
Helpful Details to Include (If Known)
- Exact accessory or module
- Part or model identification
- When recognition was lost
- Installation and seating condition
- Connector and cable condition
- Compatibility verification
- Known-good substitution result
- Restart result
- Functional test results
- Final system status
Final Thought
Confirm the correct component, inspect external installation and connections before suspecting internal electronics, prove the accessory's complete function, and escalate unresolved recognition failures appropriately.
That is successful troubleshooting.