Haemonetics Cell Saver Elite+

Centrifuge Will Not Start, Accelerate, or Maintain Speed

On this page

Asset Type

Cell Salvage System

Manufacturer

Haemonetics

Model

Cell Saver Elite+

What This Guide Helps With

Addresses centrifuge operation problems caused by loading, cover status, disposable seating, power conditions, obstruction, vibration, or service-level faults.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Continuity of Care

Stop reliance on the affected system for active blood processing. Notify the clinical team and establish an alternate blood-management plan before further troubleshooting.

Do not troubleshoot an unstable centrifuge while patient blood processing depends on it.

Expected outcome: Patient care continues safely without dependence on the malfunctioning centrifuge.

2. Confirm the Exact Centrifuge Symptom

Determine whether the centrifuge does not start at all, starts and immediately stops, accelerates slowly, produces unusual noise or vibration, or appears unable to maintain operating speed.

Document any displayed message without assuming its cause.

Expected outcome: The centrifuge failure mode is clearly characterized.

3. Verify Normal Power and Startup

Confirm the system has stable AC power, the power cord is secure, and the equipment starts without unrelated power or startup abnormalities.

Test the outlet with appropriate equipment if power quality is in question.

Expected outcome: The system has stable power and reaches its normal operating state. If power correction restores centrifuge operation, troubleshooting can stop after verification.

4. Confirm Proper Bowl and Processing-Set Installation

Inspect the bowl and disposable set for correct seating, alignment, and tubing routing. Ensure no tubing or disposable component interferes with the centrifuge area.

Do not operate the centrifuge with questionable seating.

Expected outcome: The bowl is seated securely and surrounding components are correctly installed.

5. Verify the Cover Is Fully Closed and Locked

Confirm the cover closes without obstruction and the system recognizes the locked condition.

Do not bypass the cover-lock mechanism.

Expected outcome: The system consistently recognizes the secured cover. If proper cover closure restores centrifuge operation, the issue is resolved.

6. Inspect for External Obstruction or Contamination

With the system powered down as appropriate, inspect accessible centrifuge surroundings for foreign material, dried fluid, damaged disposable components, or anything that could interfere with rotation.

Do not manually force or disassemble internal rotating assemblies.

Expected outcome: The centrifuge area is visibly clear of external obstruction.

7. Try a Known-Good Compatible Disposable

If appropriate, install a new known-good bowl and processing set to rule out deformation or improper disposable fit.

Expected outcome: The known-good disposable loads correctly and the centrifuge operates normally. If so, the original disposable was the likely external cause.

8. Observe a Nonpatient Centrifuge Start

Perform an appropriate operational check and observe startup for abnormal scraping, vibration, hesitation, or repeated stopping.

Stop the test immediately if mechanical noise, excessive vibration, odor, or other unsafe behavior occurs.

Expected outcome: The centrifuge starts and operates smoothly without abnormal mechanical behavior.

9. Verify Stable Operation Through an Appropriate Test Cycle

If the centrifuge starts normally, verify that it remains stable through an appropriate nonpatient functional sequence using manufacturer-approved procedures.

Expected outcome: The centrifuge starts, accelerates, and remains stable without interruption. Troubleshooting can stop if all applicable operational checks pass.

10. Remove From Service if Speed Control Remains Unreliable

If the centrifuge cannot start or operate consistently after external causes are ruled out, remove the Cell Saver Elite+ from service.

Label it Out of Service and send it for qualified bench evaluation.

Expected outcome: An unreliable rotating system is prevented from clinical use.

If the Problem Persists

Power, disposable loading, bowl seating, cover status, obstruction, and obvious external causes have been ruled out. The remaining fault may involve the centrifuge drive, speed sensing, mechanical alignment, control electronics, or another internal service-level condition.

The device should be:

Complete manufacturer-appropriate operational testing before return to service. Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Unusual centrifuge noise or vibration is a stop condition; do not continue operating the system in hopes that it will stabilize.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"OR staff reported that the centrifuge started briefly and then stopped before blood processing could begin."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the processing bowl incompletely seated in the centrifuge area."

Resolution

What action was taken.

Example:
"The bowl was reinstalled correctly with a verified compatible processing set, and the centrifuge completed functional testing without abnormal stopping or vibration."

Helpful Details to Include (If Known)

Final Thought

Centrifuge troubleshooting should progress from safe patient transition to power, disposable seating, cover status, and obstruction checks before internal failure is suspected. Stop when mechanical reliability cannot be demonstrated and document the result clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide