Haemonetics TEG 6s

Printer or Result Report Output Failure

On this page

Asset Type

Viscoelastic Hemostasis Analyzer

Manufacturer

Haemonetics

Model

TEG 6s

What This Guide Helps With

Addresses missing or poor report output caused by printer power, media, connection, selection, software, or external reporting-path issues.

Step-by-Step Troubleshooting

1. Protect Result Availability

If printed reports are part of the clinical workflow, confirm results remain available through another approved method while printer output is unavailable.

Do not delay critical communication solely because a printer is malfunctioning.

Expected outcome: Clinicians retain access to verified results through an alternate approved pathway.

2. Confirm the Exact Output Failure

Determine whether:

Expected outcome: The printer or report-output failure is clearly defined.

3. Verify Printer Power and Status

If an external printer is used, confirm:

Expected outcome: The printer is powered and ready. Correct simple power or ready-state issues and retest.

4. Inspect Paper or Media

Check:

Do not perform deep printer disassembly.

Expected outcome: Media is properly loaded and the printer is mechanically ready.

5. Inspect Data Connections

Check accessible USB, Ethernet, or other approved printer connection for:

Reseat connections where appropriate.

Expected outcome: The printer data connection is secure.

6. Substitute a Known-Good Cable or Printer Where Practical

Use a known-good compatible cable or approved printer comparison if available.

Expected outcome: Output resumes with the substitute component, isolating the external accessory as the cause.

7. Verify Report Selection and Output Destination

Confirm the analyzer or connected system is targeting the expected printer or report destination.

Do not change protected configuration without authorization.

Expected outcome: The intended output destination is selected.

8. Confirm Results Exist Before Troubleshooting Printing

Verify that the analyzer generated and stored or displayed the result locally.

If the result itself is missing, troubleshoot assay generation rather than printer output.

Expected outcome: A valid result is available for printing.

9. Perform a Test Report

Use an approved non-patient or previously verified result, where allowed, to confirm report generation and print output.

Check:

Expected outcome: The report prints completely and legibly. Troubleshooting can stop.

10. Escalate Persistent Output Failure

If the printer is functional, connections are verified, and the analyzer still cannot generate or route reports, escalate for software, configuration, interface, or hardware evaluation.

Expected outcome: The unresolved output issue is moved to qualified technical support without affecting patient-result integrity.

If the Problem Persists

External printer power, media, cable, output destination, and local result availability have been verified. Remaining causes may involve printer electronics, analyzer software, driver or interface configuration, network printing, or another service-level problem.

The affected equipment should be:

Verify complete and accurate report output before normal service resumes.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

A printing failure should never prevent a verified result from reaching the clinical team; use the laboratory's approved alternate reporting process.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported that completed TEG 6s results would not print to the connected report printer."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the printer data cable was loose at the printer connection."

Resolution

What action was taken.

Example:
"Reseated the cable, printed multiple verified reports successfully, confirmed complete readable output, and returned the system to normal service."

Helpful Details to Include (If Known)

Final Thought

Separate report-generation problems from printer problems, verify power and connections first, confirm the result itself is intact, and require complete readable output before closing the work order.

That is successful troubleshooting.