Haemonetics TEG 6s

Quality Control Fails or QC Result Is Out of Range

On this page

Asset Type

Viscoelastic Hemostasis Analyzer

Manufacturer

Haemonetics

Model

TEG 6s

What This Guide Helps With

Addresses failed or out-of-range quality control caused by QC material, cartridge, handling, environment, setup, or analyzer conditions.

Step-by-Step Troubleshooting

1. Protect Patient Testing

Do not release patient results from an analyzer that has not met required quality-control expectations.

Move clinical testing to another verified analyzer or approved alternate workflow until the QC issue is resolved.

Expected outcome: Patient testing is protected from an analyzer with unresolved QC performance.

2. Confirm the Exact QC Failure

Record:

Expected outcome: The QC failure is clearly characterized before corrective action begins.

3. Inspect QC Material

Verify the QC material is appropriate for the assay and has been handled according to its labeling and laboratory procedure.

Check for:

Expected outcome: QC material is suitable for use. Replace questionable material before assessing the analyzer.

4. Verify Cartridge Type and Condition

Confirm the correct cartridge is being used and inspect it for package damage, contamination, improper storage, or physical defects.

Expected outcome: The cartridge is appropriate and intact. Replace a suspect cartridge and repeat QC.

5. Review QC Preparation and Transfer

Confirm the QC material was introduced into the cartridge through the approved workflow without:

Expected outcome: QC preparation and loading are performed consistently. If corrected technique produces an acceptable result, troubleshooting can stop.

6. Repeat QC With Fresh Materials

Use fresh appropriate QC material and a known-good cartridge when available.

Avoid repeating indefinitely with the same questionable materials.

Expected outcome: QC passes when fresh verified materials are used. If so, document the material or cartridge issue.

7. Check Analyzer Environment

Verify the analyzer is:

Expected outcome: Environmental contributors are ruled out or corrected.

8. Verify Analyzer Readiness and Configuration

Confirm normal startup and verify that the intended assay and QC workflow are selected.

Do not alter protected configuration or assay parameters without authorization.

Expected outcome: The analyzer is in a normal operating state with the expected workflow selected.

9. Repeat QC and Evaluate Reproducibility

Repeat the appropriate QC process after external causes are corrected.

Expected outcome: QC meets the laboratory's approved acceptance criteria. Once required QC passes and no other abnormal behavior exists, troubleshooting can stop.

10. Escalate Repeated QC Failure

If fresh QC material and known-good cartridges repeatedly fail on one analyzer, remove it from service.

Expected outcome: The analyzer is withheld from patient testing and sent for qualified evaluation.

If the Problem Persists

QC material, cartridge condition, preparation, handling, environment, workflow selection, and basic analyzer readiness have been ruled out. Persistent QC failure may involve calibration status, internal sensing or measurement functions, software, cartridge interaction, or another service-level condition.

The device should be:

Required QC and return-to-service testing must pass before patient testing resumes.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Failed QC is a patient-safety condition, not merely a maintenance inconvenience; do not release the analyzer until required QC has been successfully repeated.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported that the TEG 6s repeatedly failed the scheduled quality-control run."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the QC material in use had been improperly stored and fresh QC material produced acceptable results."

Resolution

What action was taken.

Example:
"Removed the compromised QC material, repeated control testing with properly handled material and an appropriate cartridge, verified acceptable QC performance, and returned the analyzer to service."

Helpful Details to Include (If Known)

Final Thought

Treat failed QC as a clinical reliability issue, verify control materials and cartridges before blaming the analyzer, require successful repeat QC before release, and escalate persistent failures with clear documentation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide