Haemonetics TEG 6s

TEG Manager, LIS, or Network Communication Failure

On this page

Asset Type

Viscoelastic Hemostasis Analyzer

Manufacturer

Haemonetics

Model

TEG 6s

What This Guide Helps With

Troubleshoots lost communication between the analyzer, TEG Manager, LIS, and network infrastructure using external connectivity and configuration checks first.

Step-by-Step Troubleshooting

1. Protect the Result Reporting Workflow

If analyzer results are clinically required but electronic transmission is unavailable, use the laboratory's approved downtime or alternate reporting process.

Do not assume a result was delivered merely because testing completed locally.

Expected outcome: Results remain available to clinicians through a verified pathway while communication troubleshooting proceeds.

2. Define Where Communication Stops

Determine whether:

Expected outcome: The failed segment of the communication path is identified.

3. Verify Analyzer Network Status

Inspect accessible network indicators and the analyzer's normal connection status display, if available.

Record any displayed connection message without changing configuration.

Expected outcome: The analyzer shows either an established connection or a clearly isolated network failure.

4. Inspect Ethernet Connections

If wired networking is used, verify:

Reseat accessible connections.

Expected outcome: Physical network connections are secure. If communication resumes, verify result transmission and stop troubleshooting.

5. Substitute a Known-Good Network Cable

Use a known-good compatible cable when practical.

Expected outcome: Communication is restored with the replacement cable, confirming the original cable was defective.

6. Verify the Network Port

Coordinate with IT or network support to confirm the connected port is active and assigned as intended.

Do not make unauthorized network changes.

Expected outcome: The port is active and appropriate for the analyzer. Infrastructure issues are escalated to the responsible team.

7. Compare Other Systems

Check whether:

Expected outcome: A device-specific failure is separated from a broader server or network outage.

8. Verify Approved Communication Configuration

Review visible network or interface settings against known-good documentation or another properly configured analyzer.

Do not change IP, server, interface, or protected communication parameters without authorization.

Expected outcome: No obvious configuration mismatch is identified, or an authorized configuration correction restores communication.

9. Perform End-to-End Verification

After corrective action, verify the entire intended pathway:

Expected outcome: Communication succeeds across the complete configured path. Troubleshooting can stop.

10. Escalate Unresolved Communication Failure

If cabling, network port, server availability, and approved configuration are verified but communication remains unavailable, escalate to the appropriate Clinical Engineering, IT, LIS, or vendor support path.

Expected outcome: The problem is transferred to the team responsible for the unresolved infrastructure, interface, or application layer.

If the Problem Persists

External causes involving physical cabling, network port availability, server availability, interface path, and visible configuration have been ruled out. Remaining causes may involve internal network hardware, analyzer software, interface services, TEG Manager, LIS middleware, firewall rules, routing, or other infrastructure-level conditions.

The analyzer should be:

Complete end-to-end communication verification before return to normal workflow.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Confirm the complete path from analyzer to receiving system; a locally displayed result does not prove that the clinician or LIS received it.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported that TEG 6s results were available locally but were not appearing in TEG Manager or the LIS."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the analyzer Ethernet cable was not fully seated at the network connection."

Resolution

What action was taken.

Example:
"Reseated the network cable, verified restored connectivity and successful end-to-end result transmission, and returned the analyzer to normal service."

Helpful Details to Include (If Known)

Final Thought

Troubleshoot the communication path in layers, starting at the physical connection and ending at the receiving application, verify end-to-end delivery, and escalate at the correct infrastructure boundary.

That is successful troubleshooting.