Haemonetics TEG 6s

Unexpected Shutdown or Power Supply Failure

On this page

Asset Type

Viscoelastic Hemostasis Analyzer

Manufacturer

Haemonetics

Model

TEG 6s

What This Guide Helps With

Troubleshoots loss of power, unexpected shutdown, intermittent operation, outlet, cord, connection, accessory, and environmental causes before internal power failure is suspected.

Step-by-Step Troubleshooting

1. Protect Patient Testing and Preserve Results

If the analyzer shuts down unexpectedly, do not continue using it for patient testing until reliable power is verified.

Move clinically necessary testing to another verified analyzer or approved alternate workflow.

Expected outcome: Patient testing is transferred away from an analyzer with unreliable power.

2. Confirm the Shutdown Pattern

Determine whether:

Expected outcome: The power failure pattern is documented before components are disturbed.

3. Inspect the Power Cord and Plug

Check externally for:

Remove damaged power accessories from use.

Expected outcome: The cord and plug are intact and securely connected. Replace an approved damaged detachable cord when appropriate and retest.

4. Verify the Wall Outlet

Test the outlet using an approved method or move the analyzer to a known-good suitable outlet.

If the branch circuit or receptacle is suspect, involve Facilities or the appropriate electrical support team.

Expected outcome: Reliable facility power is confirmed. If the analyzer operates normally on another verified outlet, the analyzer may not be the cause.

5. Check the Analyzer Power Connection

Verify the power connector is fully inserted and not loose.

Gently inspect for excessive movement without stressing the connector.

Expected outcome: The connection remains stable and the analyzer does not lose power when normally positioned.

6. Remove Nonessential External Accessories

Disconnect nonessential external USB or other accessories if they can be safely removed and are not needed for the power-on test.

Expected outcome: If the analyzer becomes stable with an accessory removed, evaluate the accessory separately.

7. Inspect the Environment

Check for:

Do not open the chassis for cleaning.

Expected outcome: The analyzer is dry, adequately ventilated, and operating in a suitable environment.

8. Perform a Controlled Power-Up

Connect the analyzer to a verified power source and perform a normal startup.

Observe for:

Expected outcome: The analyzer starts normally and remains powered.

9. Perform an Extended Functional Verification

Allow the analyzer to remain powered through an appropriate observation period and approved functional checks.

Verify:

Expected outcome: Power remains stable through functional verification. Troubleshooting can stop if the original cause was external and all return-to-service requirements are satisfied.

10. Escalate Any Recurrent Power Instability

Immediately remove the analyzer from service if it:

Expected outcome: The analyzer is prevented from clinical use until qualified bench evaluation confirms reliable operation.

If the Problem Persists

External causes involving power cord, plug, outlet, connector seating, accessories, ventilation, and environmental conditions have been ruled out. Remaining causes may involve the internal power supply, power distribution, protective circuitry, thermal management, internal connections, or other service-level electronics.

The device should be:

Complete appropriate electrical-safety, startup, functional, and QC testing before return to service.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

An analyzer that has unexpectedly lost power should not be trusted for time-sensitive testing until stable operation has been demonstrated and any interrupted assay is handled appropriately.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported that the TEG 6s unexpectedly shut down during routine operation and would intermittently lose power."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the detachable power cord was loose at the analyzer and the connection would separate with normal equipment movement."

Resolution

What action was taken.

Example:
"Replaced the damaged detachable power cord with an approved known-good cord, verified stable startup and operation without further shutdowns, completed functional verification, and returned the analyzer to service."

Helpful Details to Include (If Known)

Final Thought

Treat unexpected shutdown as a reliability and patient-workflow concern, verify external power components before assuming internal failure, require stable operation before release, and escalate recurrent power problems promptly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide